Operating instructions and installation information METTLER TOLEDO MultiRange ID7-Base 2000 weighing terminal
This equipment has been tested and found to comply with the limits for a Class A digital device, pursuant to both parts of the FCC Rules and the radio interference regulations of the Canadian Department of Communications. These limits are designed to provide reasonable protection against harmful interference when the equipment is operated in a commercial environment.
Contents ID7-Base Contents Page 1 1.1 1.2 1.3 1.4 1.5 Introduction and commissioning ................................................... Safety precautions ........................................................................ Applications ................................................................................. ID7-Base weighing terminal ........................................................... Commissioning ............................................................................
Contents 2 ID7-Base 6 6.1 6.2 6.3 Application blocks ....................................................................... Syntax and formats....................................................................... TERMINAL, SCALE application blocks.............................................. INTERFACE application blocks........................................................ 7 What to do if …?......................................................................... 90 8 8.1 8.
Introduction and commissioning ID7-Base 1 Introduction and commissioning 1.1 Safety precautions ▲ Never operate the ID7-Base weighing terminal in hazardous areas; there are special scales in our product line for this purpose. ▲ Make sure that the electrical outlet for the ID7-Base weighing terminal is grounded and easily accessible to that it can be isolated quickly in emergencies. ▲ Make sure that the mains voltage at the installation location is within the range from 100 V to 240 V.
Introduction and commissioning 1.3 1.3.
Introduction and commissioning ID7-Base 1.3.2 Keypad 1 CODE A ...
Introduction and commissioning 1.3.
Introduction and commissioning ID7-Base Notes • COM1 is permanently equipped with the serial interface RS232-ID7 as standard. • Only one Alibi Memory ID7 can be installed. It has no additional external connection, and internally it occupies the space of a data port COM2 ... COM6. Alibi Memory ID7 is installed as COM4 at the factory. • A maximum of 4 cards Analog Scale-ID7, Ethernet-ID7, Profibus-DP-ID7, WLAN-ID7 or Alibi Memory-ID7 can be installed in the ID7-Base.
Introduction and commissioning 1.4.4 ID7-Base Connect ID7-Base to network CAUTION The ID7-Base weighing terminal only functions properly with mains voltages of 100 V to 240 V. ➜ Make sure that the mains voltage at the installation location lies within this range. ➜ Make sure that the mains outlet is grounded and easily accessible. Connecting ➜ Plug mains plug of ID7-Base into a mains outlet.
Introduction and commissioning ID7-Base 3. Press ZERO-SET key again. The connected weighing platform is checked. The display shows CHECK SCALE and after the test is completed SCALE IS OK. Then the ID7-Base automatically returns to normal operation. 1.5 Cleaning DANGER OF SHOCK ➜ Do not open ID7-Base weighing terminal to clean. CAUTION ➜ Make sure that unused connection sockets are covered with protective caps to protect the socket contacts from moisture and dirt. ➜ Do not use high-pressure cleaners.
Basic functions 2 ID7-Base Basic functions 2.1 Switching on and off Switch on from the standby mode ➜ Press 0N/OFF key. The display shows a weight value based on the last tare value and zero point. Switch on with restart 1. Relieve weighing platform. 2. Press ON/OFF key and hold down until METTLER TOLEDO ID7 (factory setting) or text you have specified appears in display. Then weight value 0.000 kg appears. The weighing platform is restarted.
Basic functions ID7-Base 2.3 2.3.1 Taring Manual taring 1. Place empty container on scale. 2. Press TARE key. The tare weight is saved and the weight display set to zero. The display shows the NET symbol. Notes • When the weighing platform is relieved, the saved tare weight is displayed with a negative sign. • The weighing platform only saves one tare value. 2.3.2 Automatic taring Prerequisite AUTOTARA ON must be set in the master mode, see section 4.4. ➜ Place empty container on scale.
Basic functions ID7-Base Copy tare constant The ID7-Base has 999 tare memories for frequently used tare weights programmed in the master mode. 1. Enter memory number: 1… 999. 2. Press TARE SPECIFICATION key. The display shows the NET symbol and the net weight based on the recalled tare weight. 2.3.4 Recall currently saved tare weight The saved tare weight can be recalled at any time. ➜ Enter INFO, TARE SPECIFICATION key sequence. The saved tare weight is displayed. 2.3.
Basic functions ID7-Base 2.5 Switch over weighing platform Up to 3 weighing platforms can be connected to the ID7-Base. The weighing platform currently selected is shown on the terminal. ➜ Press SCALE key. The next weighing platform is selected. – or – ➜ Enter number of weighing platform and press SCALE key. The desired weighing platform is selected.
Additional functions 3 ID7-Base Additional functions The assignment of the 6 function keys of the ID7-Base weighing terminal differs depending on the weighing task. The current assignment is shown above the function keys. With the FUNCTION CHANGE key it is possible to switch over to other function key assignments. Independent of the application software, the ID7-Base has the following additional functions: DELT DYN UNIT X 10 GROSS MODE Weighing with the DeltaTrac, see 3.1 Dynamic weighing, see 3.
Additional functions ID7-Base Application FILLING For weighing-in to a target weight with tolerance monitoring. Example: Target weight = 1.000 kg, tolerance = 1 % Target weight not reached yet Weight within the tolerance Target weight reached exactly Application CLASSIFYING To evaluate test samples as OKAY, TOO LIGHT or TOO HEAVY, based on a target weight and specified +/– tolerances. Example: Target weight = 1.
Additional functions ID7-Base 3.1.1 Enter numerically Preset DeltaTrac target values 1. Press DELT key. 2. Enter target weight and confirm with ENTER. 3. Enter tolerance in % of target weight and confirm with ENTER. Note With the FUNCTION CHANGE key you can select the weight unit for entering the DeltaTrac target values. Correct entry Copy constants ➜ With the CLEAR key the entry is corrected character by character.
Additional functions ID7-Base 3.2 Dynamic weighing With the dynamic weighing function you can weigh restless weighing samples, e.g. live animals. To do this, specify the number of weighing cycles for which the mean weight value is to be taken. 1. Set container on the weighing platform. 2. Tare weighing platform. 3. Place weighing sample in container. 4. Press DYN key and enter number of weighing cycles. Possible values: 1 … 255. 5. Start dynamic weighing with ENTER key. 6.
Additional functions ID7-Base 3.4 Working in a higher resolution Depending on the setting of the master mode block CONTROL MODE (see page 32), the weight value can be displayed in a higher resolution continuously or when called. Weight values in a higher resolution are marked with a *. Displaying weight values in higher resolution ➜ Press X 10 key. The weight value is displayed in at least a 10x higher resolution. The higher resolution is displayed until the X 10 key is pressed again.
Additional functions ID7-Base 3.7 Multiplicative tare function The multiplicative tare function is particularly suitable when pallets with identical containers are filled. If the number of containers and tare of the individual container are known, the ID7-Base weighing terminal calculates the total tare. 1. Press MULT TARE key. 2. Enter known tare weight of individual container and confirm with ENTER. 3. Enter number of containers and confirm with ENTER.
Additional functions ID7-Base 3.10 Display ID code and test weighing platform Each time the weighing platform configuration is changed the ID code counter is increased by 1. On certified weighing platforms the displayed ID code must match the ID code on the ID code sticker, otherwise the calibration is no longer valid. Display ID code ➜ Press ZERO-SET key and hold until IDENTCODE = ... appears in the display and press again. Test weighing platform ➜ Press ZERO-SET key again.
Additional functions ID7-Base Enter alphanumeric identification 1. Press one of the keys CODE A … CODE D. The functions keys are given the following assignment: ABCDE FGHIJ KLMNO PQRST UVWXY Z/-() Selection of letters A to E Selection of letters F to J Selection of letters K to O Selection of letters P to T Selection of letters U to Y Selection of letter Z and special characters 2. Select desired group of letters, e. g. press KLMNO key. 3. Select desired letter.
Additional functions ID7-Base 3.12 Recall information On the ID7-Base weighing terminal memory contents and system information can be recalled. 1. Press INFO key. Then the following function key assignment appears: DELT TARE TEXT ALIBI DATE VERS Display DeltaTrac values Display tare weight Display fixed texts and name of keys CODE A … CODE D Recall content of alibi memory, see section 2.1. This selection only appears when Alibi Memory-ID7 is installed.
Additional functions ID7-Base 3.13 Print or transfer data If a printer or computer is connected, weighing results can be printed out or transferred to the computer. In the master mode you can set the following for this purpose: • Data to be printed or transferred, • Manual or automatic data transfer, • Key which triggers printing or data transfer. Factory setting • Manual triggering with the ENTER key. • The content of the display is transferred or printed. 3.
Additional functions ID7-Base 3.15 Working with external keypad In addition to the alpha and numerical keys, the following additional scale functions can also be operated with the external AK-MFII keypad.
Additional functions ID7-Base 3.16 Working with a second display An ID1 Plus, ID3s or another ID7-... weighing terminal can be connected to the ID7-Base weighing terminal as a second display. Conditions • Interface CL 20mA-ID7 installed in passive operating mode (factory setting). • AUTO-DIR setting selected in mastermode (see page 41). • Weighing terminal is connected as second display with cable 00 504 511.
Settings in the master mode 4 ID7-Base Settings in the master mode 4.1 Overview of the master mode In the master mode you adapt the ID7-Base weighing terminal to meet your needs. Depending on the configuration, the master mode is divided into 4 or 5 master mode blocks, which are in turn divided into further blocks. SERVICE TERMINAL PAC 26 For system settings, such as entering the date and time or loading permanent texts, see section 4.3.2. To set application-specific parameters.
Settings in the master mode ID7-Base 4.2 4.2.1 Operating the master mode Enter the master mode 1. Press MODE key. If the current function key assignment does not contain MODE, change to the assignment with MODE by repeatedly pressing the FUNCTION CHANGE key. 2. Enter personal code if configured. The display shows the first master mode block TERMINAL. 4.2.
Settings in the master mode 4.2.5 ID7-Base Emergency entrance into the master mode If a personal code has been assigned for entering the master mode and you have forgotten your code, you can still enter the master mode: ➜ Enter the character sequence C, L, E, A, R as your personal code. 4.3 4.3.
Settings in the master mode ID7-Base 4.3.2 Settings in the TERMINAL master mode block FIX-TARE Save tare values protected against power failure to tare memories LOAD FIX-TARE 1. Enter memory number of FIX-TARE NO.: 1 … 999. 2. Enter tare weight for the selected memory in the displayed unit. 3. To load additional fixed tare values, repeat the first two steps. 4. End entry: Confirm FIX-TARE NO. without entry with ENTER. CLEAR ALL TARES Delete all tare memories.
Settings in the master mode ID7-Base DELTATRAC Set DeltaTrac application TYPE Select DeltaTrac application FILLING Weigh in target weight within a tolerance range (factory setting) CLASSIFYING Evaluate the test samples as good, too light or too heavy based on the target weight and tolerance CHECKWEIGHING Determine difference between target and actual weight AUTO PRINT WITHIN TOL Automatic printout when actual weight lies within the specified tolerance PRINT ONLY WITHIN TOL Printout only when a
Settings in the master mode ID7-Base KEYS A B C D Name identification keys CODE A … CODE D KEY A Identifikation data CODE A TEXT Factory setting: ARTICLE NO. DATA LENGTH Max. 30 characters possible, factory setting: 20 characters KEY B Identifikation data CODE B TEXT Factory setting: ORDER NO. DATA LENGTH Max. 30 characters possible, factory setting: 20 characters KEY C Identifikation data CODE C TEXT Factory setting: CODE NO. DATA LENGTH Max.
Settings in the master mode ID7-Base MASTER MODE START POS. Select start position for entering the master mode NORMAL Selection of the master mode blocks always begins with the TERMINAL block (factory setting). LAST POSITION When entering the master mode, the last block edited is displayed immediately. SCREEN SAVER Switch screen saver on or off WAITING TIME Enter time until screen saver is activated. Possible values: 1 ...
Settings in the master mode ID7-Base ID5 MODE Deactivating or activating downward compatibility with ID5 If ID5 MODE ON is selected, the ID7-Base is operated with downward compatibility to the ID5. Affected settings Text length of identification data Text length for keys CODE A ... D Date/time Barcode print command 18 characters max.
Settings in the master mode RESET TERMINAL ID7-Base Reset all terminal functions to the factory setting DELTATRAC MASTER MODE START POS. BIG WEIGHT DISPLAY DYNAMIC WEIGHING CONTROL MODE ID5 MODE DISPLAY DURATION ENTRY PROMPT Comment 34 Filling Normal On No printout X 10 key Off 2 / 3 seconds Off The memories are not affected by this.
Settings in the master mode ID7-Base 4.4 SCALE master mode block In the first block the weighing platform is selected: SCALE 1 … SCALE 3. The other setting possibilities are the same for all connected weighing platforms. 4.4.1 Legend Overview of the SCALE master mode block In the SCALE master mode block the following settings for the weight can be carried out: • Blocks highlighted in grey are described in detail in the following. • Factory settings are printed in bold print.
Settings in the master mode 4.4.2 ID7-Base Settings in the SCALE master mode block WEIGHING-PROC ADAPT Adapt weighing platform to weighing sample UNIVERSAL WEIGHING For solid bodies, coarse filling or checkweighing (factory setting). STATIC WEIGHING For solid bodies and weighing under extreme conditions, e. g. strong vibrations or weighing animals. FINE FILLING For liquid or powdered weighing samples.
Settings in the master mode ID7-Base RESTART Switch restart function on or off When RESTART ON is set, the zero point and tare value remain stored after the power supply is interrupted. When the weighing platform is switched on again, the terminal shows the current weight. Factory setting: RESTART OFF SECOND UNIT Select second weight unit Possible units: g, kg, lb, oz, ozt, dwt Unit Abbreviation Conversion to g Kilogram kg = 1000 g Pound lb ≈ 453.59237 g Ounce oz ≈ 28.349523125 g Troy Ounce ozt ≈ 31.
Settings in the master mode 4.5 Select the interface connection Select interface type Possible interface types ID7-Base INTERFACE master mode block ➜ Select the interface connection in the first block: COM1, COM2, COM3, COM4, COM5 or COM6. ➜ Specify the interface type for the selected interface connection COM1 … COM6. • NOT ASSIGNED When the selected interface connection is not assigned. • GA46 For connection of the GA46/GA46-W printer. Data is exchanged via the RS232-ID7 interface.
Settings in the master mode ID7-Base 4.5.1 Legend • SCALE SICS For connecting a reference scale. For this purpose an RS232ID7 interface must be installed on the selected interface connection and the reference scale must be capable of processing at least the SICS Level 0 command set. This selection no longer appears if a total of 3 scales are already connected to the ID7-Base. When SCALE SICS is selected, the following default settings are set: SICS mode, 9600 baud, 8 data bits, 1 stop bit, no parity.
Settings in the master mode 4.5.2 ID7-Base Settings in the master mode blocks RS232, RS422, RS485, CL20mA RS232, RS422, RS485, CL20mA OPERATING MODE 1:1 CONNECTION ID7-Base weighing terminal and peripheral are directly connected. BUS SLAVE For operating the ID7-Base weighing terminal in a bus system. The following parameters are set automatically for the dialog: No handshake, no continuous transmission, no transfer string, fixed string framing CRLF.
Settings in the master mode ID7-Base Set dialog mode DIALOG MODE Set dialog between ID7-Base weighing terminal and computer MMR For information on dialog mode with the MMR command set, see section 5.1. All parameters are shown on a display page and can be set there; for function key assignment, see page 42. HANDSHAKE Possible settings: • NO HANDSHAKE • CL HANDSHAKE – for additional information on the CL handshake, see page 43. • XON-XOFF PROTOCOL.
Settings in the master mode ID7-Base DIALOG MODE Set dialog between ID7-Base weighing terminal and computer TOLEDO CONTINUOUS For the continuous transmission of net and tare values to METTLER TOLEDO devices, e. g. to a second display. For a description, see section 5.2. This block does not appear with the RS485-ID7 interface. CHECKSUM ON Checksum byte active, factory setting CHECKSUM OFF Checksum byte inactive, the transfer format is shortened by 1 character.
Settings in the master mode ID7-Base CL handshake With the CL handshake 3 types of interface control are possible: Handshake in receiving direction, in transmitting direction and in both directions. After switch-on and after each interruption, the ID7-Base attempts to establish the handshake in both directions. CL handshake in receiving direction This type of CL handshake is suitable for data transmission from the ID7-Base to the computer. 1. The ID7-Base transmits SYN after switch-on. 2.
Settings in the master mode ID7-Base Set print mode PRINT MODE Configure printout on an external printer HANDSHAKE Possible settings: • NO HANDSHAKE • XON-XOFF PROTOCOL LINE LENGTH Enter number of characters per line. Possible settings: 1 … 80 characters Factory setting: 40 characters LINE FRAMING Enter ASCII character for line framing. Possible settings: ASCII 0 … 255 Factory setting: ASCII 013 010 (CRLF) REPORT TYPE Assignment of one of two possible printout formats to the configured printer.
Settings in the master mode ID7-Base Change configuration Function keys The function keys are assigned in CHANGE CONFIGURATION as follows: < > Fu ADD ↑ Display previous entry Display next entry Select function of function key F5: ADD, INS etc. ADD INS EDIT DEL PRINT Return to next highest level; changes are not saved The printout can be edited with function key F5: ADD INS EDIT DEL PRINT Adds a new entry at the end of the printout. Inserts a new entry in front of the displayed entry.
Settings in the master mode ID7-Base TYPE parameter Selection possibilities: AB Output content of an application block with or without designation TEXT Print out any desired text CHRn Insert n of any desired ASCII characters in the line, e.g. for tables; selection of character via DATA parameter LINE Blank line or separator line with any desired alphanumeric characters DB Accesses a database field. When a field is printed out, all entries of the field are listed.
Settings in the master mode ID7-Base FILL parameter PAD parameter DATA/FIELD parameter Entry of DATA parameter Show content with leading blank spaces up to maximum available length; the FILL parameter is only available for: • Designation and content separated with extra blank spaces • Content alone, left-justified • Content alone, centred Show designation and content separated with x blank spaces Possible settings: 0 ... 63 extra blank spaces.
Settings in the master mode ID7-Base Enter ASCII code for LINE and CHRn parameters 1. Open the entry mask with the EDIT key. 2. Press the +/– key and enter the ASCII code numerically. 3. Complete the numeric entry with the +/– key. 4. Complete entry with ENTER. 4.5.3 BARCODE Set barcode reader Set barcode reader TYPE DL900/DL910 DLL6000 ... Select barcode reader. When one of the barcode readers is selected, the communication and mode parameters for the selected barcode reader are automatically set.
Settings in the master mode ID7-Base 4.5.4 Configure inputs/outputs 4 I/O / RELAY BOX 8 INPUT Operate inputs internally or externally. INTERNALLY Factory setting. Additional settings: CONFIGURE INPUTS Select the desired setting for every input. Factory setting for ID7-Base: Input 1 not in use Input 2 zero setting Input 3 taring Input 4 entry (ENTER key) Input 5 ... not in use Input 8 Additional settings, only for 4 I/O: ON/OFF HIGH ACTIVE Factory setting, the ID7-Base is switched off when ON/OFF = 1.
Settings in the master mode ID7-Base 4 I/O / RELAY BOX 8 I/O TEST Testing of the function and state of the inputs and outputs of one or two connected 8-ID7 relay box(es) If an input or output is set (high), the display indicates its number. If an input or output is not set (low), the display indicates –. Set outputs Switch over the outputs with the keys 1 to 8 of the numerical keypad. Set inputs Set inputs, e. g. by connecting a supply voltage (+24 V).
Settings in the master mode ID7-Base SETPOINT MODE ON – defining set points After SETPOINT MODE ON is selected, the following input mask appears (Example): SP1: F↑ AO12 W1 1.2345 KG SP2: F↓ AO13 W2 0.
Interface description 5 ID7-Base Interface description To exchange data with a computer, the ID7-Base weighing terminal is equipped with an RS232 interface. Up to 5 additional interfaces are available as an option. The interfaces operate independently of each other, can be used simultaneously and can be adjusted individually, see section 4.5.
Interface description ID7-Base Data formats • The following symbols are used in the following command description: Xfjhiu!wbmvf Voju Ufyu`o 10 characters with sign and decimal point, right-justified (with preceding blank spaces) 3 characters, left-justified (with following blank spaces) maximum of n characters, left-justified • The string framing is mandatory, however it is not contained in the following command description! • Enter commands as ASCII characters.
Interface description ID7-Base 5.1.
Interface description ID7-Base 5.1.3 Command description Switch keypad on or off Command {R[0} {R[1} Switch on keypad Switch off keypad Response {R[B} Keypad switched on or off Comments • Factory setting: Keypad switched on. • When the keypad is switched off, the terminal cannot be operated manually.
Interface description ID7-Base Tare Command {T} Tare weighing platform: After the weighing platform stabilizes, the current weight value is saved as the tare weight and the weight display is set to zero with the weight placed on the platform. Effect as when TARE key is pressed.
Interface description ID7-Base Specify DeltaTrac target value Command Response Comments {D[Y|_|Ubshfu!xfjhiu!)xfjhiu!wbmvf*|_|Voju|_|Upmfsbodf|_|%} {D[Y} Specify DeltaTrac target value Delete DeltaTrac target value {D[B} DeltaTrac target value loaded/deleted • Observe limit values, see section 3.1.1 • Also possible: {A[W[0[2[0[..., see section 6.2 Example Command: {D[Y|_|4[.
Interface description ID7-Base Transmit data record Command Response Transmit a data record with stabilized weight values after weighing platform stabilization. Effect as if ENTER key is pressed. {S[X[I} Transmit a data record with stabilized or dynamic weight values independent of weighing platform stabilization. {S[X[I[R} Transmit data records with stabilized or dynamic weight values repeatedly independent of weighing platform stabilization.
Interface description ID7-Base Read application block Command Response Comments {A[R|Op/} Read content of application block {A[B|_|Jogpsnbujpo} Content of application block transmitted • Transmitted information is dependent on application block, see chapter 6. • Number of application block must be entered as 3 digits with preceding zeros.
Interface description ID7-Base Alphanumeric printout on GA46 printer Command Response Comments {P|_|Ufyu`59} {P|_|$|!|1|Ufyu`59} {P|_|$|!|2|Ufyu`59} {P|_|$|!|3|Ufyu`59} {P|_|$|!|A|Ufyu`59} {P|_|$|!|B|Ufyu`59} {P|_|$|!|C|Ufyu`59} {P|_} Print text as per setting Print text in small type Print text in normal type Print text in large type Print text in small type and bold print Print text in normal type and bold print Print text in large type and bold print Print blank line {P[B} Alphanumeric characters
Interface description ID7-Base Actuating digital outputs Command {W|_|Tubuvt} Switch individual digital outputs on or off {W|_|Tubuvt!2|_|Ujnf!2|_|Tubuvt!3|_|Ujnf!3|_|...|Tubuvt!5|_|Ujnf!5|_|Tubuvt!6} {W}, {W|_} Status: Response Comments Trigger time sequence of status changes of digital outputs Reset all outputs to logical 0 Time: Each output is assigned a value. The total of the values of those outputs which are to be closed is indicated as the "Status".
Interface description ID7-Base 5.1.4 Terminal messages – only with RS232, RS422 or C20mA In the dialog mode the ID7-Base weighing terminal transmits an acknowledgement to the computer each time a key is pressed. When this pressing of a key is replaced with an interface command, the acknowledgement only differs in the second character in the response format which is part of the command: Function Key Acknowledgement Set zero {Z[A} Tare {T[A}...
Interface description ID7-Base 5.2 METTLER TOLEDO continuous mode These operating modes are suitable for continuous data transmission in real time from the ID7-Base to METTLER TOLEDO devices, e. g. to a second display. The data are even transmitted when the weighing platform is moving or the gross weight = 0. Commands can also be sent to the ID7-Base weighing terminal, permitting remote control of certain keys on the terminal.
Interface description ID7-Base Status byte SB2 Bit 6 Bit 5 Bit 4 Bit 3 Bit 2 Bit 1 Bit 0 0 1 0 lb 0 Stabilization 0 Normal status 0 Positive sign 0 Gross value 1 kg 1 Movement 1 Underload/ overload 1 Negative sign 1 Net value Status byte SB3 5.2.
Interface description ID7-Base 5.3 5.3.1 METTLER TOLEDO SICS command set Communication syntax and formats Command format when transmitting weight values Identification Response format when transmitting weight values Identification Example Data formats _ Weight value String of characters for specification of command (1 … 4 characters) String of characters for specification of response (1 ... 2 char.) 1 … 10 characters _ Status 1 char. _ Unit Framing 1 ...
Interface description ID7-Base 5.3.
Interface description ID7-Base 5.3.3 Command description Transmit SICS commands Command Response {I[0} Transmit SICS commands {I[0|_|B} {I[0|_|0|_|"J1"} {I[0|_|0|_|"J2"} ... {I[0|_|1|_|"E"} ... {I[0|_|2|_|"TY"} ... {I[0|_|3|_|"BS"} ...
Interface description ID7-Base Transmit scale software version Command Response Example {I[3} Transmit software version from weighing terminal and weighing platform(s) {I[3|_|A|_|"ufyu"} {I[3|_|A|_|"JQ84.1.1311J!J[16.1.1412!J[21.1.1332"} Transmit serial number Command Response Example Comment {I[4} Transmit serial number of weighing terminal {I[4|_|A|_|"ufyu"} {I[4|_|A|_|"2345678"} The response to I4 appears automatically following switch-on and after the Reset command (@).
Interface description ID7-Base Set to zero Command Response {Z} Set gross weight display to zero after weighing platform comes to a standstill, effect as when ZERO-SET key is pressed {Z|_|A} {Z|_|I} Weighing platform set to zero Command cannot be executed: e.g.
Interface description ID7-Base Keyboard monitoring Command {K[_[1} {K[_[1} {K[_[3} {K[_[4} Response {K|_|A} {K|_|I} {K|_|L} When a key is pressed, execute the function, but do not transmit anything (factory setting) When a key is pressed, do not execute the function and do not transmit anything When a key is pressed, do not execute the function, but transmit the key code {K[_[C[_[x} or, when the key is pressed longer, transmit {K[_[R[_[x} and {K[_[C[_[x} When a key is pressed, execute the function an
Interface description ID7-Base Response Comments Function codes x The function codes are dependent on the command transmitted.
Interface description ID7-Base Transmit stable weight values repeatedly depending on a weight change Command {S[R|_|Fydvstjpo!xfjhiu!)xfjhiu!wbmvf*|_|Voju} {S[R|} Response After a weight change greater than the specified excursion weight, transmit alternately the next stable weight value and a dynamic weight value depending on the specified excursion. If no excursion weight is entered, the weight change must be at least 12.5 % of the last stable weight value, however at least 30 d.
Interface description ID7-Base Tare immediately Command Response {T[I} Tare weighing platform immediately.
Interface description ID7-Base Delete tare weight Command Response {T[A[C} Delete tare weight. {T[A[C|_|A} {T[A[C|_|I} Weighing platform tared with the specified weight Command not carried out Transmit data record Command Response After the weighing platform comes to a standstill, transmit a data record with stable weight values. Effect as when ENTER key is pressed. {S[X[I} Transmit a data record with stable or dynamic weight values, regardless of whether the weighing platform is at a standstill.
Interface description ID7-Base Switch keyboard on or off Command Response Comments {R[0} {R[1} Switch on keyboard Switch off keyboard {R[0|_|A} Keyboard switched on {R[1|_|A} Keyboard switched off • Factory setting: Keyboard switched on. • When the keyboard is switched off, the terminal cannot be manually operated.
Interface description ID7-Base Write application block Command Response Comments {A[W|_|Op/|_|Jogpsnbujpo} {A[W|_|Op/} {A[W|_|Op/|_} Write application block Reset application block Delete application block {A[W|_|A} {A[W|_|I} {A[W|_|L} Application block written Application block not present Application block cannot be written • The information to be entered is dependent on the target block, see chapter 6. • Deleting and resetting have the same effect.
Interface description ID7-Base Print text or barcode with GA46 printer Command Response Comments {P|_|Ufyu`59} Print text as per setting {P|_|$|!|1|Ufyu`59} Print text in small print {P|_|$|!|2|Ufyu`59} Print text in normal print {P|_|$|!|3|Ufyu`59} Print text in large print {P|_|$|!|A|Ufyu`59} Print text in small type and bold print {P|_|$|!|B|Ufyu`59} Print text in normal type and bold print {P|_|$|!|C|Ufyu`59} Print text in large type and bold print {P|_|$|#|1|Ufyu`31-!cbsdpef.
Interface description ID7-Base Actuating digital outputs Command {W|_|Tubuvt} Switch individual digital outputs on or off {W|_|Tubuvt!2|_|Ujnf!2|_|Tubuvt!3|_|Ujnf!3|_|...|Tubuvt!5|_|Ujnf!5|_|Tubuvt!6} {W}, {W|_} Status: Response Comments Trigger time sequence of status changes of digital outputs Reset all outputs to logical 0 Time: Each output is assigned a value. The total of the values of those outputs which are to be closed is indicated as the "Status".
Interface description ID7-Base 5.3.4 Error messages Error messages always consist of 2 characters and a string limit. The string limit can be defined in the master mode (section 4.5.2). {E[T} Transmission error The terminal transmits a transmission error for errors in the received bit sequence, e.g. parity error, missing stop bit. {E[S} Syntax error The terminal transmits a syntax error when it cannot process the received characters, e.g. command not present.
Application blocks 6 ID7-Base Application blocks Application blocks are internal information memories in which weighing data, calculated quantities, configuration data or character sequences entered with the keypad are stored. The content of the application blocks can be read out or written to with a computer. When the GA46 printer is connected, the assignment of the application blocks can be printed out, see operating instructions for the GA46 printer. 6.
Application blocks ID7-Base Note If an application block is not in use, the weighing terminal transmits the corresponding number of blank spaces in place of the data. For example, when Tare Memory 1 is not in use, the weighing terminal transmits the following response: {A[B|_|_[_[_[_[_[_[_[_[_[_|_|_[_[_} (MMR) resp. {A[W|_|A|_|_[_[_[_[_[_[_[_[_[_|_|_[_[_} (SICS) 6.1.
Application blocks ID7-Base 6.1.3 Data formats • In the following description of the application blocks the following data formats are used: Xfjhiu!wbmvf Voju Ovncfs`o Ufyu`o 10 digits with sign and decimal point, right-justified (with preceding blank space) 3 characters, left-justified (with following blank spaces) Number, n digits, right-justified (with preceding blank spaces) maximum of n characters If the SICS command set is used, "Text" must always be placed in inverted commas.
Application blocks ID7-Base 6.2 TERMINAL, SCALE application blocks No.
Application blocks ID7-Base No. Content Format 019 Date and time Response: Write: {A[B|_|_[_[_[_[_[_[D[D[/[M[M[/[Y[Y|_[_| {A[B|_|_[_[_[_[_[_[h[h[:[m[m[:[s[s} Europe {A[B|_|_[_[_[_[_[_[M[M[/[D[D[/[Y[Y|_[_| {A[B|_|_[_[_[A/P[M[_[h[h[:[m[m[:[s[s} USA {A[W|0[1[9|_|D[D[/[M[M[/[Y[Y|$|$| {A[W|0[1[9|_|h[h[:[m[m[:[s[s} Europe {A[W|0[1[9|_|M[M[/[D[D[/[Y[Y|$|$| {A[W|0[1[9|_|h[h[:[m[m[:[s[s[_[A/P[M} USA Date: instead of "/" also "."; Time: instead of ":" also "/" or ".
Application blocks ID7-Base No. Content Format 091 Barcode EAN 28, EAN 128 Response: FBO!39: {A[B|_|FBO!39|_|_|FBO!239!12|_|_|FBO!239!421|_|_| {A[B|_|FBO!239!441} |2[8[Bsujdmf[Difdl!ejhju[Xfjhiu| Article: 4-digit Article No. from memory Code A Check digit: 1-digit, calculated by ID7-Base for the weight Weight: 5-digit positive weight value with 3 decimal places between 00.000 kg - 99.
Application blocks ID7-Base No.
Application blocks ID7-Base 6.3 INTERFACE application blocks Application blocks are reserved for the possible interface connections. These application blocks can only be read and written to when an ...-ID7 interface is actually installed on the interface connection concerned. 6.3.1 Serial interfaces No.
Application blocks ID7-Base 6.3.2 Digital inputs/outputs The following application blocks are only available when interface 4 I/O-ID7 is installed on COM5/COM6 or interface RS485-ID7 and relay box 8-ID7 is installed on COM6. When the weighing terminal checks the outputs, the blocks concerned cannot be written to, and the {E[L} error message appears. No. Content Format 706 Digital outputs 1 COM5/COM6 Response: Write: {A[B|_|9.qmbdf!cjobsz!wbmvf} * {A[W|7[0[6|_|9.
Application blocks ID7-Base No. Content Format 724 Set point 1 Response: Write: Note: {A[B|_|Tfu!qpjou!)Ufyu`3*!!!!!!!!!!!|_[_| |A[x[x[x[_[y[y[y[.[z[z|_[_| |Tdbmf!)Ufyu`4*!!!!!!!!!!!!!!!!!!!!|_[_| |Tfu!qpjou!wbmvf!)xfjhiu!wbmvf*!!!!|_[_} {A[W|7[2[x|_|Tfu!qpjou!uzqf!)Ufyu`3*!!!!!!!!!!!|$[$| |A[x[x[x[_[y[y[y[.
What to do if …? 7 ID7-Base What to do if …? Error / Display Possible causes Remedy Display is dark • No mains voltage ➜ Check mains • Terminal switched off ➜ Switch on terminal • Power cord not connected ➜ Plug in power plug • Brief malfunction ➜ Switch terminal off and on again • Load plate not in place ➜ Apply load plate • Preload not applied ➜ Apply preload • Weighing range dropped below ➜ Set zero • Weighing range exceeded ➜ Relieve weighing platform • Weighing platform locked
What to do if …? ID7-Base Error / Display Possible causes Remedy SCALE NO.
What to do if …? ID7-Base Error / Display Possible causes Remedy DISPLAY MODE • Weighing cell defective ➜ Contact METTLER TOLEDO Customer Service • 2 weighing platforms with same scale ➜ Contact METTLER TOLEDO Customer number connected Service 92 Operating instructions and installation information 22004109D 04/10
Technical data and accessories ID7-Base 8 Technical data and accessories 8.
Technical data and accessories ID7-Base Weighing functions Tare compensation At the press of a button or automatically, up to maximum load (subtractive) Tare target value • For single-range scales over entire weighing range (subtractive) • For multi-range scales depending on national calibration regulations • 999 stored tare memories, protected against power failure Tare calculation Tare addition, tare multiplication, sub-tare Tare indicator NET lights up with saved tare weight DeltaTrac • Analog
Technical data and accessories ID7-Base Interface RS232-ID7 Interface type Voltage interface as per EIA RS232C/DIN 66020 (CCITT V.24/V.
Technical data and accessories ID7-Base Dimensions Dimensions in mm 96 Operating instructions and installation information 22004109D 04/10
Technical data and accessories ID7-Base 8.2 Accessories Applications Order No.
Technical data and accessories ID7-Base Serial data interfaces Order No. CL20mA-ID7 CL 20 mA interface 22 001 084 Accessories for CL20mA-ID7 CL cable, 3 m Mating connector, 7-pin Second-display cable CL20mA-ID7 – ID1 Plus/ID3s/ID7, 10 m Extension cable for second display, 10-pin, 10 m Adapter cable PE / CL, 0.
Technical data and accessories ID7-Base Digital/analog interface Order No. Analog Output-ID7 Digital/analog output 0 – 10 V, 0 – 20 mA or 4 – 20 mA 22 001 090 Accessories for Analog Output-ID7 Cable for Analog Output-ID7, 5-pin, 3 m Mating connector, 5-pin 00 204 930 00 205 538 Alibi memory Alibi Memory-ID7 Order No. Paperless archiving of certification-relevant weighing data Network connections 22 001 663 Order No.
Technical data and accessories ID7-Base External keypad Order No. AK-MFII Compact, alphanumeric membrane keypad for connection to the standard-equipment 5-pin MFII circular connector Housing of all nickel chromium steel, protection type IP65 Dimensions (W x D x H): 380 mm x 158 mm x 30 mm Cable approx. 1 m 00 505 490 Accessories for AK-MFII Keypad-terminal adapter 00 208 047 Other accessories Order No.
Appendix ID7-Base 9 Appendix 9.
Appendix ID7-Base 9.2 Key numbers All keys of the ID7-Base are assigned numbers so that they may be addressed via the interface. 102 Key Number Key Number Key 0 0 SCALE key 19 Key 1 1 ZERO-SET key 20 ... ...
Appendix ID7-Base 9.3 Opening/closing ID7-Base weighing terminal CAUTION ➜ The ID7-Base weighing terminal may only be opened by authorized personnel! ➜ Always pull the mains plug before opening the unit. Opening 1. Unscrew the screws on the underside of cover. 2. Set down the cover toward the front. Closing 1. Lay the device on the cover and loosely secure with the 3 screws. 2. Press the device into the cover so that all 2 locking springs engage. 3. Screw on the cover with 3 screws.
Index ID7-Base 10 Index A Accessories 97 Acoustic signal 60, 75 Additional functions 14 Additive tare function 19 Alibi Memory-ID7 7, 38, 99 Alphanumeric identification 21 Alphanumeric keypad 24, 100 Alphanumeric printout 60, 77 Analog Output-ID7 6, 99 Application blocks 59, 75, 76, 80 Applications 3 ASCII characters 53, 101 Automatic continuous transmission 41 Automatic taring 11, 36, 94 Autozero 10, 36 B Barcode 23, 38, 48, 60 Basic functions 10 Big Weight Display 2, 4, 32, 93 Bus Slave 40, 53 C Certifi
22004109D Subject to technical changes © Mettler-Toledo (Albstadt) GmbH 04/10 Mettler-Toledo (Albstadt) GmbH D-72458 Albstadt Tel. ++49-7431-14 0, Fax ++49-7431-14 232 Internet: http://www.mt.