Operating instructions
Table Of Contents
- Operating instructions
- Operating instructions in a nutshell
- Contents
- 1 Getting to know L/L-S balances line
- 2 Startup
- 3 Weighing
- 4 Menu
- 4.1 Overview
- 4.2 Menu operation
- 4.3 Description of menu options
- 4.3.1 Reset or recording of balance settings (1st menu option “RESET”)
- 4.3.2 Functions (2nd menu option / see Section 5 for their use)
- 4.3.3 Weighing mode (3rd menu option)
- 4.3.4 Weighing unit 1 (4th menu option “UNIT 1”)
- 4.3.5 Weighing unit 2 (5th menu option “UNIT 2”)
- 4.3.6 Autozero (6th menu option / see overview and notes in Section 4.1)
- 4.3.7 Auto shut off
- 4.3.8 Peripheral unit (8th menu option / see overview and notes in Section 4.1)
- 4.3.9 Send mode (9th menu option / see overview and notes in Section 4.1)
- 4.3.10 Send format (10th menu option / see overview and notes in Section 4.1)
- 4.3.11 Baud rate (11th menu option / see overview and notes in Section 4.1)
- 4.3.12 Bit/Parity (12th menu option / see overview and notes in Section 4.1)
- 4.3.13 Handshake (13th menu option / see overview and notes in Section 4.1)
- 5 Functions
- 6 Technical data, options, optional equipment
- 7 Appendix

Startup
9
English
CAL
CAL donE
2000.00
------
0.00
------
0.00 g
▼
▼
▼
Cal
Cal
Cal
Cal
Cal
▼
2.4 Adjusting (calibration)
To obtain accurate weighing results, the balance must be adjusted to match the gravitational acceleration at its location.
Adjusting is necessary
• before the balance is used for the first time
• at regular intervals during weighing service
• after a change of location
To obtain accurate results, the balance must be left switched on for 60 minutes to
reach operating temperature before starting the adjustment procedure.
➞ Have required adjusting weight ready.
➞ Unload weighing pan.
➞ Press and hold the «Cal» key down until “CAL” appears in the display. Release
key.
The required adjustment weight value flashes in the display.
➞ Place adjustment weight in centre of pan. The balance adjusts itself auto-
matically.
➞ When “0.00 g” flashes, remove adjustment weight.
The adjusting is finished when the message “CAL done” appears briefly in the
display, followed by “0.00 g”. The balance is again in weighing mode and ready
for operation.
Notes
• Certified PL-S models cannot be adjusted by the user, because of weights and
measures legislation.
• This adjustment procedure can be terminated at any time with the «C» (“Cancel”)
key. The balance reverts to weighing mode.