Error/Error message Cause Rectification
Overload Remove sample from weighing pan, zero
again (tare).
Underload Check whether weighing pan is
positioned properly.
No stability
in taring or adjusting (calibration) Wait for stability before pressing key.
when reference weight for piece counting Ensure more stable ambient conditions.
is placed on pan Remove weighing pan and clean if
necessary
Wrong adjustment weight on pan or none at all Place required adjustment weight in
centre of pan.
Reference weight (Piece counting, Percent Increase reference weight.
weighing, Plus-minus weighing) too small
Internal fault Contact METTLER TOLEDO customer
service.
Wrong weighing pan or pan missing or Place correct pan or empty pan on
not empty balance.
Adjustment aborted with the «C» key
No display
AC adapter not plugged in Check AC power supply.
Plug AC adapter into power supply.
Batteries discharged (only with compact Replace batteries; if using rechargeables
models) connect instrument to AC power supply.
7.3 Connecting L/L-S balances to other METTLER TOLEDO devices
Device Connecting cable settings/
Remarks
Titrators:
DL31, 36, 38 RS9RS9 (m/f) send continous
DL50, 53, 55, 58 11101051
MTCom-Bus 310 RS9RSopen (m/-)
21900640
SQC 14 RS9RS9 (m/f)
(statistical quality control) 11101051
Spider (industrial scale) RS9RS9 (m/m)
Viper BC (industrial scale) 21252588
LC-PVolume (pipette calibration) RS9RS9 (m/f) AX balances recommended
11101051
LC-PCalc RS9RS9 (m/f)
11101051
LC-P45 (application printer) RS9RS9 (m/f)
11101051
GA42 (report printer) RS9RS9 (m/f)
11101051
íååååì
7.2 What if ...?
î____ï
Error 1
Error 2
Error 3
0.000
g
Abort
Error 4