Operating instructions

Table Of Contents
Technical data, options, optional equipment
26
The MT-SICS commands listed below is a selected list of available commands. For additional commands and further information
please refer to the Reference Manual MT-SICS for B-S/L/L-S balances 11780447 downloadable from the Internet under
www.mt.com/pl or www.mt.com/al.
S – Send stable weight value
Command S Send the current stable net weight value.
SI – Send value immediately
Command SI Send the current net weight value, irrespective of balance stability.
SIR – Send weight value immediately and repeat
Command SIR Send the net weight values repeatedly, irrespective of balance stability.
Z – Zero
Command Z Zero the balance.
@ – Reset
Command @ Resets the balance to the condition found after switching on, but without a zero setting
being performed.
SR – Send weight value on weight change (Send and Repeat)
Command SR Send the current stable weight value and then send continuously the stable weight
value after every weight change.
The weight change must be at least 12.5 % of the last stable weight value,
minimum = 30d.
ST – Send stable weight after pressing ± (transfer) key
Command ST Inquiry of actual status of the ST function.
SU – Send stable weight value with currently displayed unit
Command SU As the S command, but with the currently displayed unit.