Operating instructions
Table Of Contents
- Operating instructions
- Operating instructions in a nutshell
- Contents
- 1 Getting to know L/L-S balances line
- 2 Startup
- 3 Weighing
- 4 Menu
- 4.1 Overview
- 4.2 Menu operation
- 4.3 Description of menu options
- 4.3.1 Reset or recording of balance settings (1st menu option “RESET”)
- 4.3.2 Functions (2nd menu option / see Section 5 for their use)
- 4.3.3 Weighing mode (3rd menu option)
- 4.3.4 Weighing unit 1 (4th menu option “UNIT 1”)
- 4.3.5 Weighing unit 2 (5th menu option “UNIT 2”)
- 4.3.6 Autozero (6th menu option / see overview and notes in Section 4.1)
- 4.3.7 Auto shut off
- 4.3.8 Peripheral unit (8th menu option / see overview and notes in Section 4.1)
- 4.3.9 Send mode (9th menu option / see overview and notes in Section 4.1)
- 4.3.10 Send format (10th menu option / see overview and notes in Section 4.1)
- 4.3.11 Baud rate (11th menu option / see overview and notes in Section 4.1)
- 4.3.12 Bit/Parity (12th menu option / see overview and notes in Section 4.1)
- 4.3.13 Handshake (13th menu option / see overview and notes in Section 4.1)
- 5 Functions
- 6 Technical data, options, optional equipment
- 7 Appendix

Functions
21
English
Requirement
Different weight units must be activated in the menu for unit 1 and unit 2 (Section 4).
Requirement
The function “F FAC M” or “F FAC d” must be activated in the menu (Section 4).
#
#
±±
±±
±
1/10d
F
±±
±±
±
““
““
“
1/10d
““
““
“
F FAC Ÿ≈
1.0000
0.9999
#
…
0.9500
#
…
StEP≠
#
0.01
#
0.02
#
0.05
#
17.00
#
64.71 g
64710 mg
““
““
“
▼
17.89 g
5.6 Switching weight units
Entering the free factor and/or the display increments
➞ Hold the «F» key down until “F FAC M” or “F FAC d” appears in the display.
➞ Press the «
““
““
“» key to select “FAC M” / “FAC d” or “noFAC M” / “noFAC d”
(Function deactivated).
➞ Release the key. Either the factor 1 appears as default value or the factor that was
saved most recently.
This value can now be changed:
➞ Pressing the «
““
““
“» key increases the factor.
Pressing the «1/10d» key reduces the factor.
Pressing the key once changes the value by one increment. If the key is held down,
the value changes increasingly rapidly.
Confirm the selected factor with the «
±±
±±
±» key (it will not be saved
automatically). “StEP=” appears in the display, and the program changes auto-
matically to allow the display increments to be entered. The smallest possible
display increment appears as default value, or the last value that was saved.
➞ This value can be changed in the same way as for the free factor (see above).
Confirm the selected display increment with the «
±±
±±
±» key (it will not be saved
automatically).
The appropriate calculation is then made using the weight on the pan in grams and
the selected factor, the result being displayed with the selected display increment. No
units are displayed, the symbol “#” being displayed instead. The calculation is
always based on the weight in grams.
Note
• If you only want to change the display increments, set the free factor
at exactly 1.
Toggling between displaying the calculated value and the measured weight
Place the sample on the weighing pan. The appropriate calculation is then made
using the weight of the sample and the selected factor, the result being displayed
with the selected display increment.
Press the «
““
““
“» key. The weight is displayed (in unit 1, and if the key is pressed
again in unit 2, provided that this option is activated).
➞ Press the «
““
““
“» key again to return to the calculated value.
➞ The «
““
““
“» key can be used at any time to toggle between the two weighing
units selected in the menu (“UNIT 1” and “UNIT 2”).
Notes
• Switching between weight units may be blocked with certified balances,
depending on national weights and measures legislation.
• This function is not available with dynamic weighing.