Operating instructions

Table Of Contents
Functions
19
English
t ≠ 3 ”
á
t ≠ 5 ”
á
0.00 g
F
±±
±±
±
±±
±±
±
0.00 g
0.00 g
á
• 257.35 g
á
-- 3 --
á
-- 1 --
á
260.33 g
á
*
Dynamic weighing with manual start (F dYn M)
The «
» key can be used select the dynamic weighing. The display shows the
symbol á.
Load weighing sample.
Start weighing with the «
±±
±±
±» key.
During the weighing time, a count down runs in the display.
Read off result.
The result of the dynamic weighing is displayed with (= calculated value) and
remains in the display until the weighing sample is removed from the weighing
pan or the container.
Notes
The weighing cycle with the same weighing sample can be restarted with the «
±±
±±
±» key.
The «
» key can be used to switch between dynamic weighing and normal weighing.
For weighing goods below 5 g the weighing must be started manually with the «
±±
±±
±» key, even for dynamic weighing with
automatic start.
Changing the weighing time
Press and hold the «F» key, until t = 3" appears in the display.
Repeatedly press the «
» key, until the desired weighing time appears.
Possible values are 3", 5", 10", 20", 1", 2".
election with the «
±±
±±
±» key briefly to confirm or by automatic acceptance after 3
seconds.