Operating instructions
Table Of Contents
- Operating instructions
- Operating instructions in a nutshell
- Contents
- 1 Getting to know L/L-S balances line
- 2 Startup
- 3 Weighing
- 4 Menu
- 4.1 Overview
- 4.2 Menu operation
- 4.3 Description of menu options
- 4.3.1 Reset or recording of balance settings (1st menu option “RESET”)
- 4.3.2 Functions (2nd menu option / see Section 5 for their use)
- 4.3.3 Weighing mode (3rd menu option)
- 4.3.4 Weighing unit 1 (4th menu option “UNIT 1”)
- 4.3.5 Weighing unit 2 (5th menu option “UNIT 2”)
- 4.3.6 Autozero (6th menu option / see overview and notes in Section 4.1)
- 4.3.7 Auto shut off
- 4.3.8 Peripheral unit (8th menu option / see overview and notes in Section 4.1)
- 4.3.9 Send mode (9th menu option / see overview and notes in Section 4.1)
- 4.3.10 Send format (10th menu option / see overview and notes in Section 4.1)
- 4.3.11 Baud rate (11th menu option / see overview and notes in Section 4.1)
- 4.3.12 Bit/Parity (12th menu option / see overview and notes in Section 4.1)
- 4.3.13 Handshake (13th menu option / see overview and notes in Section 4.1)
- 5 Functions
- 6 Technical data, options, optional equipment
- 7 Appendix

Functions
19
English
t ≠ 3 ”
á
t ≠ 5 ”
á
0.00 g
F
““
““
“
±±
±±
±
±±
±±
±
““
““
“
0.00 g
0.00 g
á
• 257.35 g
á
-- 3 --
á
-- 1 --
á
260.33 g
á
▼
*
▼
…
Dynamic weighing with manual start (F dYn M)
➞ The «
““
““
“» key can be used select the dynamic weighing. The display shows the
symbol á.
➞ Load weighing sample.
➞ Start weighing with the «
±±
±±
±» key.
During the weighing time, a “count down” runs in the display.
➞ Read off result.
The result of the dynamic weighing is displayed with ✱ (= calculated value) and
remains in the display until the weighing sample is removed from the weighing
pan or the container.
Notes
• The weighing cycle with the same weighing sample can be restarted with the «
±±
±±
±» key.
• The «
““
““
“» key can be used to switch between dynamic weighing and normal weighing.
• For weighing goods below 5 g the weighing must be started manually with the «
±±
±±
±» key, even for dynamic weighing with
automatic start.
Changing the weighing time
➞ Press and hold the «F» key, until “t = 3"” appears in the display.
➞ Repeatedly press the «
““
““
“» key, until the desired weighing time appears.
Possible values are 3", 5", 10", 20", 1", 2".
➞ election with the «
±±
±±
±» key briefly to confirm or by automatic acceptance after 3
seconds.