Operating instructions
Table Of Contents
- Operating instructions
- Operating instructions in a nutshell
- Contents
- 1 Getting to know L/L-S balances line
- 2 Startup
- 3 Weighing
- 4 Menu
- 4.1 Overview
- 4.2 Menu operation
- 4.3 Description of menu options
- 4.3.1 Reset or recording of balance settings (1st menu option “RESET”)
- 4.3.2 Functions (2nd menu option / see Section 5 for their use)
- 4.3.3 Weighing mode (3rd menu option)
- 4.3.4 Weighing unit 1 (4th menu option “UNIT 1”)
- 4.3.5 Weighing unit 2 (5th menu option “UNIT 2”)
- 4.3.6 Autozero (6th menu option / see overview and notes in Section 4.1)
- 4.3.7 Auto shut off
- 4.3.8 Peripheral unit (8th menu option / see overview and notes in Section 4.1)
- 4.3.9 Send mode (9th menu option / see overview and notes in Section 4.1)
- 4.3.10 Send format (10th menu option / see overview and notes in Section 4.1)
- 4.3.11 Baud rate (11th menu option / see overview and notes in Section 4.1)
- 4.3.12 Bit/Parity (12th menu option / see overview and notes in Section 4.1)
- 4.3.13 Handshake (13th menu option / see overview and notes in Section 4.1)
- 5 Functions
- 6 Technical data, options, optional equipment
- 7 Appendix

Menu
15
English
S.SICS
““
““
“
S.PŸ≈
bd 2400
““
““
“
bd 4800
…
““
““
“
bd 19200
S.oFF
““
““
“
S.Stb
S.Cont
S.Auto
S.ALL
4.3.9 Send mode (9th menu option / see overview and notes in Section 4.1)
Note: This menu option is only available if the “Host” setting was selected in the 8th menu option (Peripheral unit)!
It specifies how a value is transferred to a peripheral device.
S.oFF Send mode switched off.
S.Stb The next possible stable value will be transferred after the «
±±
±±
±» key
has been pressed.
S.Cont All values are transferred automatically.
S.Auto Only stable values are transferred automatically.
S.All The current value is transferred after the «
±±
±±
±» key has been pressed.
4.3.10 Send format (10th menu option / see overview and notes in Section 4.1)
Note: This menu option is only available if the “S.oFF” setting was not selected in the 9th menu option (“Send mode”)!
It sets the data transfer format.
“S. SICS”: The MT-SICS data transfer formats are used. Please refer to the
“Reference Manual MT-SICS B-S/L/L-S balances 11780447”,
available from your METTLER TOLEDO dealer or downloaded from the
Internet (www.mt.com/pl or www.mt.com/al see “Support”). More
Information please find in the Section 6.3.
“S. PM”*: The following PM balance data transfer formats are used:
S.Stb: 1.67890 g
S.Cont: S 1.67890 g SD 1.39110 g
S.Auto: S 1.67890 g
S.All: 1.67890 g
D 1.39110 g
* unidirectional, no MT-SICS commands are accepted.
4.3.11 Baud rate (11th menu option / see overview and notes in Section 4.1)
Note: This menu option is only available if the “Printer” or “Host” setting was selected in the 8th menu option (Peripheral unit)!
The baud rate (data transfer rate) determines the speed of transmission via the serial
interface. The unit is the baud (bd) = 1 bit/second.
The following settings are available: 600 bd, 1200 bd, 2400 bd, 4800 bd, 9600 bd
and 19200 bd.
For problem-free data transmission the sending and receiving devices must be set at
the same value.