Operating instructions

Table Of Contents
Menu
15
English
S.SICS
S.PŸ≈
bd 2400
bd 4800
bd 19200
S.oFF
S.Stb
S.Cont
S.Auto
S.ALL
4.3.9 Send mode (9th menu option / see overview and notes in Section 4.1)
Note: This menu option is only available if the Host setting was selected in the 8th menu option (Peripheral unit)!
It specifies how a value is transferred to a peripheral device.
S.oFF Send mode switched off.
S.Stb The next possible stable value will be transferred after the «
±±
±±
±» key
has been pressed.
S.Cont All values are transferred automatically.
S.Auto Only stable values are transferred automatically.
S.All The current value is transferred after the «
±±
±±
±» key has been pressed.
4.3.10 Send format (10th menu option / see overview and notes in Section 4.1)
Note: This menu option is only available if the S.oFF setting was not selected in the 9th menu option (Send mode)!
It sets the data transfer format.
S. SICS: The MT-SICS data transfer formats are used. Please refer to the
Reference Manual MT-SICS B-S/L/L-S balances 11780447,
available from your METTLER TOLEDO dealer or downloaded from the
Internet (www.mt.com/pl or www.mt.com/al see Support). More
Information please find in the Section 6.3.
S. PM*: The following PM balance data transfer formats are used:
S.Stb: 1.67890 g
S.Cont: S 1.67890 g SD 1.39110 g
S.Auto: S 1.67890 g
S.All: 1.67890 g
D 1.39110 g
* unidirectional, no MT-SICS commands are accepted.
4.3.11 Baud rate (11th menu option / see overview and notes in Section 4.1)
Note: This menu option is only available if the Printer or Host setting was selected in the 8th menu option (Peripheral unit)!
The baud rate (data transfer rate) determines the speed of transmission via the serial
interface. The unit is the baud (bd) = 1 bit/second.
The following settings are available: 600 bd, 1200 bd, 2400 bd, 4800 bd, 9600 bd
and 19200 bd.
For problem-free data transmission the sending and receiving devices must be set at
the same value.