Operating instructions

Table Of Contents
Menu
14
4.3.5 Weighing unit 2 (5th menu option “UNIT 2”)
If it is required to show the weighing result in weighing mode in an additional unit by pressing the «
» key, the desired second
weighing unit can be selected in this menu option. The same weighing units are available as under UNIT 1, with the exception
of the tael units (H tl, S tl and t tl).
4.3.6 Autozero (6th menu option / see overview and notes in Section 4.1)
This menu option allows you to switch the automatic zero correction on or off.
Autozero switched on
The zero point is automatically corrected (e.g. if drift occurs or the weighing pan
becomes dirty). Certified balances, however, have a fixed zero point.
Autozero switched off
The zero point is not automatically corrected. This setting is advantageous for special
applications (e.g. evaporation measurements).
4.3.7 Auto shut off
If the automatic shut off function is activated, the balance automatically switches itself off after a selected period of inactivity (i.e.
with no key being pressed or changes of weight occurring):
ÅZEro
no ÅZEro
A.OFF 5é
A.OFF -
A.OFF 30”
A.OFF 2é
A.OFF 10é
PrintEr
Host
z.dISPLAY
A.OFF 10 Automatic shutoff after 10 minutes inactivity
A.OFF - Automatic shutoff not activated
A.OFF 30 Automatic shutoff after 30 seconds inactivity
A.OFF 2 Automatic shutoff after 2 minutes inactivity
A.OFF 5 Automatic shutoff after 5 minutes inactivity
4.3.8 Peripheral unit (8th menu option / see overview and notes in Section 4.1)
Peripheral devices can only be connected if the balance has been equipped with the optional RS232C interface. The balance
automatically saves the appropriate settings (Sections 4.3.9 4.3.13) for every peripheral device.
Printer Connected to a printer.
Host Connection to any desired peripheral device.
Aux. display Connection of an optional auxiliary display unit (communications
parameters cannot be selected).