Operating instructions
METTLER TOLEDO DE40/50/51 Density Meter 91
Specifications
Specifications
Type and model name Digital density meter DE40, DE50, DE51
Measuring method Oscillating U-tube
Density DE40 DE50, DE51
Measuring range d 0.0001 to 3 g/cm
3
0.00001 to 3 g/cm
3
Limit of error d 1 x 10
-4
2 x 10
-5
Repeatability d 1 x 10
-4
1 x 10
-5
Resolution d 1 x 10
-4
1 x 10
-5
Temperature
Control Peltier
Range T 4 – 90 °C 4 – 70 °C
Precision T 0.05 °C (10 to 30 °C) 0.02 °C (10 to 30°)
T 0.1 °C ( 4 to 90 °C) 0.05 (4 to 70 °C)
Viscosity compensation Memory Card (option) DE50: not possible
DE51: standard
Sample required min. 1.2 mL (manual injection with syringe)
min. 2 mL (automated operation with optional pump)
Typical meas. times 1 to 4 minutes (manual operation)
2 to 10 minutes (programmed automated operation)
Display LCD (Liquid crystal display): 7 line display, backlit
External I/O 1 RS-232C (for computer and combination)
1 RS-232C (for external printer)
1 special serial interface (for sample changer)
1 Memory Card (PCIMCIA)
Ambient condition 1) Temperature: 5 ~ 35 °C
2) Humidity: below 85 % RH (no condensation)
Power source AC 100-240 V, 50/60 Hz
Power consumption 200 W
Dimension 270 mm (W) x 400 mm (D) x 410 mm (H)
Weight 15 kg