Operating instructions

72 METTLER TOLEDO DE40/50/51 Density Meter
13.2.1 Select a concentration table: ALCOHOL, BRIX, CONCENTRATIONS
Press the <CALC.PARA> key.
Select “1.Result”. Confirm with
<ENTER>.
Select “Brix” (Alcohol,
Concentration, according to your
application). Confirm with
<ENTER>.
Press <ENTER> (confirm Execute).
Press <METHOD>.
The method list now should show
you the selected method with the
corresponding application.
13.2.2 Select a temperature table: API
Press the <CALC.PARA> key.
Select “2.Temperature
Compensation”. Confirm with
<ENTER>.
Select “Oil Table-1” (Oil Table-2 or
Oil Table-3, according to your
application) with the key. See
also section 14.1. Confirm with
<ENTER>.
Press <RESET>.
Application Memory Card
Method 1 <Calc.Parameter>
0.Density
1.Result
2.Temperature Compensation
3.Decimals
Method 1 <Result>
Result :<Brix >
Exit :[Execute]
< Method List >
No. Name Temp.(°C) Result .
0. Method-0 20.0 d
1. ORANGE 20.0 Brix
2. Method-2 20.0 d
3. Method-3 20.0 d
4. Method-4 20.0 d
Method 1 <Calc.Parameter>
0.Density
1.Result
2.Temperature Compensation
3.Decimals
Method 1 <Temp.Comp.>
Temp.Comp : Off On
Comp.Temp. :[20.00]°C
Temp.Formula :<Oil Table-1>