Operating instructions

METTLER TOLEDO DE40/50/51 Density Meter 71
Wait until the installation has been
done.
Press <ENTER>.
The asterisk (*) on the left of the
application (BRIX) shows that this
application has been downloaded
and removed from the Memory
Card. This means that the Memory
Card is empty, but can still be used
to retrieve the application from the
instrument (uninstall table).
13.2 How to select the application
The application has been downloaded to the instrument. You need now to define which
method should be dedicated to this application.
Press <METHOD>.
Select the method you want to
dedicate to this application with the
<UP> and <DOWN> keys. Confirm
with <ENTER>.
Press the <MEAS.PARA> key.
Set the measurement parameters
according to your application (see
section 8).
There are now three different selections, depending on the application:
1. Select concentration table: ALCOHOL, BRIX, CONCENTRATIONS.
2. Select temperature table: API.
3. Select viscosity correction: see section 6.2.
< Install Table >
Now Install ...
Table install is complete
[ OK ]
< Install Table >
Table Name Type Temp. Data.
*Brix Conc. 20.00°C 85
Exit Install
Application Memory Card
< Method List >
No. Name Temp.(°C) Result .
0. Method-0 20.0 d
1. Method-1 20.0 d
2. Method-2 20.0 d
3. Method-3 20.0 d
4. Method-4 20.0 d
Method 1 <Meas.Parameter>
Method Name :[ORANGE]
Measurement Temp.:[ 20.00]°C
Stability (0-2) :[1]
Limit Time :[ 600]s
Sequence :Off On