Operating instructions

70 METTLER TOLEDO DE40/50/51 Density Meter
13. Application Memory Cards: BRIX, ALCOHOL, API, VISCOSITY
CORRECTION, USER TABLES
These Application Memory Cards are FLASH cards with a capacity of 2 MB. The cards are
used to download special applications to your instrument. Once the application has been
downloaded to the instrument it is removed from the Application Memory Card. In other
words the Memory Card is empty, but can still be used to retrieve the application from the
instrument, for example to transfer it to another instrument. Therefore make sure to keep the
empty Application Memory Card in a safe place.
Note: The function "Memory Clear" "All Parameter" (in Function 9.5) doesn’t clear the
downloaded application.
13.1 How to download an application (Install table)
This section will show you how to download an application. Following Application Memory
Cards are available:
• ALCOHOL Alcohol concentration (Ethanol) (ME-51324000)
• BRIX Sugar concentration (Sucrose) (ME-51324001)
• API Petroleum products (ME-51324002)
• VISCOSITY Viscosity correction (ME-51324003)
• CONCENTRATION TABLES (ask your local representative)
Insert the Application Memory Card into the slot on the front left part of the instrument.
Press the <FUNC> key.
Select “8.CARD utility”. Confirm with
<ENTER>.
Select “6.Install Table”. Confirm with
<ENTER>.
Note: The following shows the BRIX application. For other applications just replace BRIX
with your application (Alcohol, API, Viscosity correction, Concentration).
Select the application to load with
the <UP> and <DOWN> keys.
Select “Install” with the −> key.
Confirm with <ENTER>.
Application Memory Card
< Function >
0.Data File 5.Set Check
1.Sample File 6.Periodic
2.Method File 7.Auto Start
3.Changer 8.CARD utility
4.Factor 9.Memory Clear
< CARD utility >
0.Data delete 5.Table Capacity
1.Sample delete 6.Install Table
2.Method delete 7.Uninstall Table
3.Format
4.CARD Information
< Install Table >
Table Name Type Temp. Data.
Brix Conc. 20.00°C 85
Exit Install