Operating instructions

METTLER TOLEDO DE40/50/51 Density Meter 7
Safety measures
2. Safety measures
Measures for your safety
Ensure that you plug the power cable supplied into a socket which is
grounded! In the absence of grounding, a technical fault could be lethal.
Switch the instrument off and disconnect the power cable if you change
blown fuses! An electric shock could be lethal.
Never work in a hazardous area! The housing of the instruments is not gas
tight (explosion hazard through spark formation, corrosion by gas diffusion
into the instruments).
Measure readily inflammable, toxic and corrosive substances in a fume
hood and comply with the usual lab regulations!
Measures for operational safety
Check the set operating voltage before you switch on the instrument! The
instrument will be damaged if the operating voltage does not match the line
voltage.
Do not clean the measurement tube with concentrated sodium hydroxide
(caustic soda, NaOH) or hydrofluoric acid (HF)! Both substances chemi-
cally attack the measurement tube.
Use only fuses of the specified type if you need to change them!
Exclude the following environmental influences:
powerful vibrations.
direct sunlight.
high atmospheric humidity.
temperatures below 5 °C and above 35 °C.
• powerful electric or magnetic fields which influence the power supply
through large load fluctuations!