Operating instructions
METTLER TOLEDO DE40/50/51 Density Meter 59
< Setup >
0.Interface 5.International
1.Date & Time 6.Lock
2.Name 7.Beep
3.Serial/Version 8.Temperature
4.LCD Contrast
< Temperature >
0.Calibration
1.Adjust
11.9 Temperature calibration (Setup 8)
Press the <SETUP> key.
Select "8.Temperature".
Confirm with <ENTER>.
Select either "1.Adjust" (single point)
or "0.Calibration" (multiple point over
the hole temperature range).
Confirm with <ENTER>.
Adjust:
Performs a one-point adjustment at the measuring temperature (correcting the offset).
Select this option when the difference between Set Temperature and Cell Temperature is
less than 0.5 °C (in equilibrium).
Calibration:
Performs a multiple point adjustment, in steps of 5 °C, over the entire temperature range.
Select this option when the difference between Set Temperature and Cell Temperature is
0.5 °C or higher (in equilibrium). This temperature calibration takes approximately two hours!
This temperature calibration will adjust the internal thermistors to the certified thermistor in
the measuring cell over the whole temperature range.
Setup