Operating instructions

44 METTLER TOLEDO DE40/50/51 Density Meter
After selecting <Statis.> and con-
firmation with <ENTER>, you can
select the filter to process only the
desired results: result, method no,
method name, high sample no.
Select the desired filter to be "On"
using the key.
Now you select whether the
statistics results are to be printed
alone (Data Print "Off") or together
with a list of all the measured values
(Data Print "On").
The statistics is now shown on the
display.
To print the results of the statistics
calculation press the <PRINT> key.
Select Exit : [Execute] and confirm
with <ENTER> to exit.
10.1.2 Recalculation
This function allows to recalculate single measured values. Before you do this you have to
change the calculation formula (or calculation table) in the respective method to the desired
one (like density to BRIX, shown in this example). When you now enter the recalculation
menu the result will be shown directly according to the new calculation setting.
Select the data to be recalculated
with the <UP> and <DOWN> keys.
Select Recalc. with the and
keys and confirm with <ENTER>.
Functions
< Data File >
No. M.Name Sample Result .
0 Method-0 01-01 d 0.99820
1 Method-0 01-02 d 0.99819
2 Method-1 02-01 Br 10.81
3 Method-1 02-02 Br 10.82
Statis. Recalc. Load Save Exit
< Statistics >
Result :<Brix >
Method No. :<Off>
Method Name :<Off>
High Sample No. :<Off>
Data Print :<Off >
Statistics :[Execute]
< Statistics Result >
Results : 4
Mean : 10.82 %
SD : 0.06 %
RSD : 0.54 %
Exit :[Execute]
< Data File >
No. M.Name Sample Result .
11 Method-5 01-01 d 1.03981
12 Method-5 01-02 d 1.03980
13 Method-6 02-01 Br 10.81
14 Method-6 02-02 Br 10.82
Statis. Recalc. Load Save Exit
<Statistics>
Date : 29/06/1998 17:29
Sample No.(High) : 01
Method No. : 2
Results: 4
Mean : 10.82 %
SD : 0.06 %
RSD : 0.54 %
Name : Quality Lab 1
----------------