Operating instructions
METTLER TOLEDO DE40/50/51 Density Meter 31
8.2 Measuring temperature
This option is used to set the measuring temperature.
Enter your measuring temperature
(from 4 °C to 70 °C, 4 °C to 90 °C
for DE40) and confirm with
<ENTER>.
Note: After changing the set temperature make sure to allow enough time for the instru-
ment to warm up.
To increase the temperature: To decrease the temperature:
from to change time from to change time
10 °C ⇒ 20°C +10 °C 30 min 20 °C ⇒ 10°C-10°C 1 hour
20 °C ⇒ 30°C +10 °C 30 min 30 °C ⇒ 20°C-10°C 1 hour
20 °C ⇒ 70°C +50 °C 1 hour 70 °C ⇒ 20°C-50°C 2 hours
8.3 Measurement stability
This option is used to set the measurement stability, which allows to shorten the measuring
time. See section 3.2 "Measurement principle" for more details.
Enter your stability (0, 1 or 2) and confirm with <ENTER>.
0: Maximum stability. The measured signal (T-value) has to be stable before the density is
accepted. Longest measuring time (Time-0).
1: High stability. The final T-value is extrapolated, and the density is calculated. Medium
measuring time. ( Time-1 is approximately 1/2 of Time-0) .
2: Low stability. The final T-value is extrapolated from less values, and the density is
calculated. Lowest measuring time. ( Time-2 is approximately
1
/
4
of Time-0).
Note: Stability is fixed to zero when viscosity correction is active (section 6.2).
Method 0 < Meas.Parameter > VA
Method Name :[Method-0]
Measurement Temp.:[ 20.00]°C
Stability :[1]
Limit Time :[ 600]s
Sequence : Off On
Method 0 < Meas.Parameter > VA
Method Name :[Method-0]
Measurement Temp.:[ 20.00]°C
Stability :[1]
Limit Time :[ 600]s
Sequence : Off On
Measuring parameters