Operating instructions
22 METTLER TOLEDO DL36
Setup
8.7 Selecting entry of the sample identification
You can choose whether you enter the sample identification before or after the titration or not
at all.
Key selection Display/Entry
Setup
Setup [1-8]? 7
Enter
ID entry:before ↑
Enter
Setup [1-8]?
Description
"ID entry" flashes on the display.
Select between before, after and off.
off: no inquiry of the sample identification during the titra-
tion, but the identification last entered will be automat-
ically recorded.
8.8 Initializing default parameters
You can delete all entries you have made for the method: The default parameters will then be
reentered by the DL36 (see Section 7.1). These include
• the stirring speed (default value = 4)
• the display of the amount of moisture in mg titrated up to this point (default value for
catholyte and anolyte = 0)
the defined limit values for anolyte and catholyte (A = 1000 mg, C = 300 mg)
the message indicating these limit values have been exceeded does not appear: "Alarm
set:off" (see Section 8.1)
• date and time (default values = 95/06/01 and 00:00, see Section 8.2)
• the entry of the weight and the sample number (default setting: entry before the titration,
see Sections 8.6 and 8.7).
Use the following procedure to set the default parameters:
Key selection Display/Entry
Setup
Setup [1-8]? 8
Enter
Memory clr. :off ↑
Select
Memory clr. :on ↑
Enter
Setup [1-8]?
Description
"Memory clr." flashes in the display.
The initialization is not possible.
The initialization is possible:
The confirmation is accompanied by an audio signal.
Notice: Immediately switch the instrument off, wait five sec-
onds then switch it on again! If you do not do this,
undefined values will be entered for all parameters!
After the instrument has been switched on, "SRAM Init." is displayed: the default parameters
are initialized.
– Press any key: "DL36 V1.0" appears followed by "<Run>".