Operating instructions

METTLER TOLEDO DL36 21
Notes
1. For all balances you must select the default setting with unidirectional transmission mode
"Send Cont.".
2. If you attach a balance using the LC-RS9 cable, you must set positions 7, 3 and 4
(switches left, middle and right).
3. Computer
If you attach a computer, you must select the configuration of the RS232C interface.
Key selection Display/Entry
Setup
Setup [1-8]? 5
Enter
RS232C
Enter
Baudrate: 4800
Enter
Parity: none
Enter
Data bits: 8
Enter
Stop bits: 1
Enter
Soft HS: off
Enter
Setup [1-8]?
Description
"Option" flashes on the display.
If "Printer" or "Balance" is displayed, use Select to select
"RS232C".
Select the baud rate: 9600, 300, 600, 1200, 2400 or 4800.
Select the parity: even, odd or none.
Select the data bits: 7 or 8.
Select the stop bits: 2 or 1.
Select software handshake: on or off.
You can use the installed computer to transfer data in both directions in a specific format. A
detailed description of the communication between the DL36 and a computer can be found
in a separate set of instructions.
8.6 Selecting weight entry
You can choose whether you enter the weight before or after the titration.
Key selection Display/Entry
Setup
Setup [1-8]? 6
Enter
Wt.entry:before
Select
Wt.entry:after
Enter
Setup [1-8]?
Description
"Wt.entry" flashes in the display.
You enter the weight before the titration.
You enter the weight after the titration.
Setup