Operating instructions

18 METTLER TOLEDO DL36
Setup
Key selection Display/Entry (example)
Setup
Setup [1-8]? 1
Enter
A.Capa. [mg]? 61
Enter
C.Capa. [mg]? 61
Enter
Alarm set: off
Select
Alarm set: on
Enter
A.Alarm[mg]? 1000 ..850
Enter
C.Alarm[mg]? 300 ...250
Enter
Setup [1-8]?
Description
"Reagent capacity" flashes on the display.
Display of the summed electrolytic current for the
anolyte – converted to mg H
2
0. When you change
the anolyte, enter 0.
Display of the summed electrolytic current for the
catholyte – converted to mg H
2
0. When you change
the catholyte, enter 0.
You are not informed when the defined capacity
limits are exceeded (see below).
You are informed when the defined capacity limits
are exceeded (see below). After a titration, the dis-
play then shows "A.Capacity Over".
If necessary, change the default value for the capac-
ity limit of the anolyte.
If necessary, change the default value for the capac-
ity limit of the catholyte.
Press Esc if you wish to have the current status
displayed.
8.2 Entering the date and time
The DL36 has an internal clock. When the instrument is delivered, default values are set for
the date and time (95/06/01 and 00:00) from which they will continously update. You should
change both values to ensure correct data are reported.
Key selection Display/Entry (example)
Setup
Setup [1-8]? 2
Enter
YY MM DD? 950728 ..... 960718
Enter
HH MM? 1051 ............. 1346
Enter
Setup [1-8]?
Description
"Date ? Time" flashes on the display.
Enter year, month and day.
Enter hours and minutes.
Press Esc if you wish to have the current status
displayed.