Operating instructions

14 METTLER TOLEDO DL36
7.3.2 Result parameters
This parameter group comprises three parts:
1. the calculation parameters (Calculation)
2. the selection of the report (Report)
3. the parameters for the recalculation (Recalculation).
The tables shown follow this subdivision.
1. Calculation parameters
Key selection Display/Entry (example)
Result
Calculation
Enter
Unit: µg
Select
Unit: mg
Select
Unit: ppm
Select
Unit: %
Enter
Weight: var
Select
Weight: fix
Enter
Weight? 5.0000 ..... 0.1
Enter
Drift comp: off
Select
Drift comp: manu
Select
Drift comp: auto
Enter
Blank? 0 .............. 2.5
Enter
Factor? 1.0
Enter
Calculation
Description
Select the result unit for the calculation of the mois-
ture content (the absolute µg content is always re-
corded). The choice of unit automatically deter-mines
the formula used for the calculation (see next page).
Select the entry of the weight [g]:
var: variable weight; inquiry of the weight for every
titration.
fix: fixed value; the weight for the samples being
titrated does not change, the inquiry does not
appear again in the titration. If you need only the
absolute moisture content without weight value,
enter 0 (zero).
Select the drift compensation for the titration:
off: the drift value calculated before the titration is
neither incorporated in the calculation nor re-
corded.
manu: you enter a drift value which is incorporated in
the calculation.
auto: the drift value calculated before the titration is
automatically incorporated in the calculation.
Enter a blank value [µg], e.g. if you have dissolved a
sample in a solvent whose moisture content you
have determined beforehand. This blank value is
taken into account in the calculation (see next page).
Enter a factor used to multiply the result, e.g. if you
wish to convert the result to an amount of sample
which you can not titrate.
To titrate, press Esc, to select the report parameters,
press Select.
The method