Operating instructions
25
7.3 What if...?
Error/error message
Cause
Overload
Underload
No or wrong adjusting weight on pan
➞ Wait for stability before pressing key.
➞ Ensure more stable ambient conditions.
➞ Remove sample from weighing pan, rezero (tare).
Rectification
➞ Check whether weighing pan is positioned properly.
➞ Place required adjusting weight in center of pan.
➞ Increase reference weight or piece number.
Reference weight or reference number too small
No stability
• in taring or adjusting (calibration)
• when reference weight for piece counting or percent
weighing is placed on the pan
Wrong or no weighing pan
➞ Mount correct weighing pan.
Abort of the adjustment using the
→
0/T
←
key.
Balance/scale software not sufficiently up-to-date for operation
with LC-B interface.
➞ To change the balance/scale software 299702 (Order
No.␣ 600150), ask your METTLER TOLEDO dealer.
Error 1
ñ------ó
Error 2
Error 3
ò------ô
Abort
0
Error 5