Operating instructions
20
AB54 AB104 AB204 PB153 PB303 PB303 PB302 PB602 PB1502
DeltaRange
Readability 0.1 mg 0.1 mg 0.1 mg 0.001 g 0.001 g 0.001 g*/0.01 g 0.01 g 0.01 g 0.01 g
Maximum capacity 51 g 101 g 210 g 151 g 310 g 60 g*/310 g 310 g 610 g 1510 g
Repeatability (s) 0.1 mg 0.1 mg 0.1 mg 0.001 g 0.001 g 0.001 g*/0.005 g 0.01 g 0.01 g 0.01 g
Linearity –/+ 0.2 mg 0.2 mg 0.3 mg 0.002 g 0.003 g 0.002 g*/0.02 g 0.02 g 0.02 g 0.02 g
Adjusting weight 50 g 100 g 200 g 100 g
1)
200 g
1)
200 g
1)
200 g
1)
500 g
1)
1000 g
1)
Adjusting weight with certified balances 50 g 100 g 200 g 150 g
1)
300 g
1)
300 g
1)
–
2)
600 g
1)
1500 g
1)
Overall dimensions
(WxDxH) in mm 190x290x265 190x290x339 190x290x204 190x290x62
Weighing pan in mm ø 80 ø 80 ø 100 180x166
Max. height above weighing pan 159 mm 232 mm 118 mm
Net weight (with packaging) 4.6 kg (7.3 kg) 5 kg (8.1 kg) 3 kg (5 kg) 2.5 kg (4.5 kg)
6 Technical data and optional equipment
6.1 Technical data
Standard equipment of the B balances
• Protective cover
• AC adapter to national codes
(Power supply: 115 V/230 V, –20/+15%, 50/60
Hz, 65/30 mA, 6 VA)
(Power supply balance: 8–14.5 V, 50/60 Hz or
9.5–20 V = 1.5 VA)
• Device for weighing below the balance for AB and
PB balances. With SB balances, this device is an
optional accessory.
• Adjusting weight with AB balances
• RS232C interface
Degree of protection
• Protects against dust and water.
• Pollution degree: 2
• Overvoltage category: II
Ambient conditions
The technical data are valid under the following
ambient conditions:
• Ambient temperature 10 °C…30 °C
• Relative atmospheric 15 %…85 %,
humidity noncondensing
• Height above sea level up to 4000 m
The operability is assured between ambient tem-
peratures 5–40°C.
* Fine range values (DeltaRange)
1)
Optional equipment
2)