Operating instructions

16
5.1 Piece counting
Requirement
The function "F count" must be activated in the menu, see section 4.
Place empty container on the balance and tare: press
0/T
briefly.
Setting the reference
A reference weight (reference) must first be entered for piece counting.
Add reference parts to container, possible reference numbers are 10, 20, 30, 50, 100 and 5.
Press and hold
F
until "SEt ... PCS" is displayed.
Press repeatedly until the display matches the loaded reference number.
Press briefly to confirm reference or automatic acceptance after 2 seconds. The current piece
number (PCS = pieces) is displayed.
Notes
The current reference weight remains stored until the reference is reset or the power supply is interrupted.
Counting / switching
Add weighing sample to the container and read piece number.
Press briefly, the weight is displayed.
Return to display of the piece number: press again.
5 Functions
kg
12.07
sET 10
F
PCS
20
PCS
sET 20
PCS
kg
0.00
147.25
g
244
PCS
244
PCS