4.2.4 Selecting unit 1
Depending on requirements, the balance can operate with the following units:
Unit Conversion factor Comments
g gram factory setting
kg kilogram 1 kg = 1000 g not with 1 mg balances
lb pound 1 lb 453.59237 g not with 0.1 mg balances
oz ounce 1 oz 28.349523125 g
ozt troy ounce 1 ozt 31.1034768 g
GN grain 1 GN 0.06479891 g not with 1 g balances
dwt pennyweight 1 dwt 1.555173843 g
ct carat 1 ct = 0.2 g
mg milligram 1 mg = 0.001 g only with 0.1 mg and 1 mg balances
mo momme 1 mo 3.749999953 g
m Mesghal 1 m 4.6083162 g
H tl Hong Kong taels 1 H tl 37.42900 g selectable only in unit 2
S tl Singapore taels 1 S tl 37.799366256 g selectable only in unit 2
t tl Taiwan taels 1 t tl 37.499995313 g selectable only in unit 2
4.2.5 Selecting unit 2
If the weighing results should be shown in a different unit in the weighing mode by pressing
(see section␣ 5.4), the appropriate unit must be selected in the menu.
4.2.6 A. oFF – automatic shutdown
The automatic shutdown extends the operating time in line-independent operation with the PP-B10 PowerPack
considerably.
When the automatic shutdown is activated, the balance switches itself off if no weighing has been performed
during the specified time. With the PowerPack the balance is then off, without the PowerPack it is on standby.
A. oFF – no automatic shutdown (factory setting)
A. oFF 2 automatic shutdown after 2 minutes
A. oFF 5 automatic shutdown after 5 minutes
A. oFF 10 automatic shutdown after 10 minutes