Operating instructions
8
2.4 Adjusting (calibration)
To obtain accurate weighing results, the balance must be matched to the acceleration due to gravity at its
location.
Adjusting is necessary
• before the balance is used for the first time,
• at regular intervals during weighing operations,
• after a change in location.
Procedure
To obtain accurate results, before adjusting the balance must be switched on for 20–30 minutes so that the
operating temperature will be reached.
➞ Have required adjusting weight ready (section 6.1).
➞ Unload weighing pan.
➞ Press and hold
Cal/Menu
until "CAL" appears in the display, release key.
The required adjusting weight value flashes in the display.
➞ Place adjusting weight in center of pan. The balance adjusts itself.
➞ When "0.00" flashes, remove adjusting weight.
The adjusting (calibration) is finished when "0.00 g" appears in the display. The balance is again in the
weighing mode and ready for operation.
Note
• Depending on national certification specifications, the adjusting may be locked with certified balances after
the installation.
• The adjustment can be terminated at any time using the
→
0/T
←
key. The following message appears:
Abort.
CAL
------
2000.00
------
0.00
------
CAL donE
Cal/Menu
0.00
g