6.3 What if...?
Error/error message
Cause
Overload
Underload
➞ Wait for stability before pressing key.
➞ Ensure more stable ambient conditions.
➞ Remove sample from weighing pan, rezero (tare).
If overload appears when balance switched on,
recalibrate (readjust) balance.
Rectification
➞ Remove sample from weighing pan, rezero.
➞ Check whether weighing pan is positioned properly.
If underload appears when balance switched on,
recalibrate (readjust) balance.
No stability in taring or calibration (adjusting)
No or wrong calibration weight on pan
➞ Place required calibration weight in center of pan.
➞ Mount correct weighing pan.
Wrong or missing weighing pan
Abort of the adjustment using the
#
key.
Balance/scale software not sufficiently up-to-date for operation
with LC-B interface.
➞ To change the balance/scale software 299702 (Order
No. 600150), ask your METTLER TOLEDO dealer.
Error 5
Error 1
ñ------ó
ò------ô
Error 2