Operating instructions

11
3.5 Switching of weight units
All B balances can display weight values in different weight units.
With school balances (College-B), the fixed units gram (g) and newton (N*) are available.
* As the balance is an instrument to determine mass, when school balances are used to determine force a
small measurement error arises. It depends on the geographic location as a constant conversion factor 1
kg = 9.80665 N is used.
With carat and gold scales (CB, GB), 3 different units can be displayed in the weighing mode. These are freely
selectable in the menu.
The factory setting differs according to the country and the model.
Switching between unit 1, unit 2 and unit 3
Requirement
With CB and GB scales, different units must be activated in the menu (section 4) for Unit 1, 2 and 3.
ct
110.00
g
22.00
dwt
14.14
Press again, the display switches to the 3rd weight unit.
Note
With certified balances/scales, the unit selection can be blocked following installation if required by national
legislation.
Press
, the display switches to the 2nd weight unit.
^