Operating instructions
36
Cause
7.2 What if ...?
Error/Error message
Overload ➞ Remove sample from weighing pan, zero again (tare).
Underload ➞ Check whether weighing pan is positioned properly.
No stability
• in taring or adjusting (calibration)
• when reference weight for piece counting is placed on pan
➞ Wait for stability before pressing key.
➞ Ensure more stable ambient conditions.
➞ Remove weighing pan and clean if necessary
Wrong adjustment weight on pan or none at all ➞ Place required adjustment weight in center of pan.
Reference number for piece counting too small ➞ Increase number of reference pieces.
íååååì
î____ï
Error 1
Error 2
Error 3
0.000
g
Abort
Wrong weighing pan or pan missing
Error 4
Internal fault ➞ Contact METTLER TOLEDO customer service.
Error 6
No standard calibration ➞ Contact METTLER TOLEDO customer service.
➞ Mount correct weighing pan.
Adjustment aborted with the «C» key
Rectification