Operating instructions

26
Dynamic weighing with manual start (F dYn M)
The «» key can be used select the dynamic weighing. The display shows the symbol ,.
Load weighing sample.
Start weighing with the «±» key.
During the weighing time, a count down runs in the display.
Read off result.
The result of the dynamic weighing is displayed with (= calculated value) and remains in the display
until the weighing sample is removed from the weighing pan or the container.
Notes
The weighing cycle with the same weighing sample can be restarted with the «±» key.
The «» key can be used to switch between dynamic weighing and normal weighing.
For weighing goods below 5 g the weighing must be started manually with the «±» key, even for
dynamic weighing with automatic start.
Changing the weighing time
Press and hold the «F» key, until t = 3" appears in the display.
Repeatedly press the «» key, until the desired weighing time appears.
Possible values are 3", 5", 10", 20", 1", 2".
Selection with the «±» key briefly to confirm or by automatic acceptance after 3 seconds.
Note
The set weighing time remains stored until it is reset or the power supply fails.