Operating instructions

23
5.1 Piece counting
5 Functions
Requirement
The function F count must be activated in the menu (Section 4).
Place empty container on the balance and tare by briefly pressing the «#» key.
Setting the reference: a reference weight must first be entered for piece counting:
Add a number of reference pieces to container. Possible numbers are 5, 10, 20, 50, 100 and no (this
setting deactivates the piece counting function).
Caution: Take into account minimum values: min. reference weight = 10d (10 digits), min. piece
weight = 1d (1 digit)!
Note: 1 digit corresponds to 1 display increment.
Hold the «F» key down until SEt ... PCS is displayed.
Repeatedly press the «» key until the display equals the number of reference pieces entered.
Confirm the number of reference pieces with the «±» key (or wait 7 seconds, in which case the number
is adopted automatically). The current number of pieces (PCS = pieces) is displayed.
Note: The current reference weight remains stored until the reference setting is changed or the power supply
is interrupted.
Switching between piece count and weight display
Add weighing sample to the container and read off number of pieces.
Press the «» key. The weight is displayed.
Return to the piece count display by pressing the «» key again.
0.00 g
#
F
SEt 10
PCS
SEt 20
PCS
±
20
PCS
240
PCS
144.81 g
M
E
T
T
L
E
R
T
O
L
E
D
O
F
C
a
l
/
M
e
n
u
C
On
o
ff
O/
T
M
E
T
T
L
E
R
T
O
L
E
D
O
F
C
a
l
/
M
e
n
u
C
On
o
f
f
O/
T
M
E
T
T
L
E
R
T
O
L
E
D
O
F
C
a
l
/
M
e
n
u
C
On
of
f
O/T