Operating instructions

21
4.3.11 Send mode (11th menu option)
Note: This menu option is only available if you have selected the setting Host in the 10th menu option
(Peripheral unit)!
At this menu option you specify how data is transmitted to a peripheral device.
S. oFF Send mode switched off
S. Stb The next possible stable value will be transferred after the «±» key has been triggered.
S. Cont All values are transferred automatically.
4.3.12 Send format (12th menu option)
Note: This menu option is only available if you have selected the setting S. Stb or S. Cont at the 11th menu
option (Send mode)!
At this menu option you specify the data transfer format.
S. SICS: The MT-SICS data transfer formats are used. Please refer to the MT-SICS Reference Manual
Basic-S balances 11780447 available from your METTLER TOLEDO dealer or download from the
Internet (www.mt.com/AB-S). More Information please find in the Section 6.3.
S. PM*: The following PM balance data transfer formats are used:
S. Stb: /////1.67890/g S. Cont: S////1.67890/g
SD///1.39110/g
* unidirectional, no MT-SICS commands are accepted.
4.3.13 Baud rate (13th menu option)
Note: This menu option is only available if you have selected the setting Host in the 10th menu option
(Peripheral unit)!
The baud rate (data transfer rate) determines the speed of transmission via the serial interface. The unit is the
baud (bd) = 1 bit/second.
The following settings are available: 600 bd, 1200 bd, 2400 bd, 4800 bd, 9600 bd and 19200 bd.
For problem-free data transmission the sending and receiving devices must be set at the same value.
S. oFF
S. Stb
S. Cont
S. SICS
S. PŸ≈
bd 2400
bd 4800
bd 19200