Operating instructions

10
2.4 Adjusting (calibration)
To obtain accurate weighing results, the balance must be adjusted to match the gravitational acceleration at its
location.
Adjusting is necessary
before the balance is used for the first time
at regular intervals during weighing service
after a change of location
Procedure
To obtain accurate results, the balance must be connected to the power supply for 30 minutes (AB-S analytical
balances 60 minutes) in order to reach operating temperature before adjusting.
Analytical balances (AB-S), Precision balances (PB-S certified)
Adjusting with internal weight
To carry out this operation, in the second menu option (Adjustment) select CAL int (=factory setting)
(Section 4.1).
Unload weighing pan
Press and hold the «Cal/Menu» key down until CAL appears in the display, then release key.
The balance adjusts itself automatically.
The adjusting is finished when the message CAL done appears briefly in the display, followed by
0.0000 g. The balance is again in weighing mode and ready for operation.
Analytical balances (AB-S)
Adusting with external weight
To carry out this operation, in the second menu option (Adjustment) select CAL E (=factory setting)
(Section 4.1).
Then proceed as for precision balances.
Cal/Menu
CAL E
------
à
CAL donE
0.0000 g
à
à
Cal/Menu
CAL int
Cal