Technical data

9
04/12
2 Setting up
2.1 Scope of delivery
f Check if the delivery is complete:
Scale
Stainless steel weighing pan
Power supply, stored in the power supply storage compartment
under the scale
Operating instructions
Tower with mounting material, tower version only
2.2 Requirements for scale’s location
The right location is critical to weighing accuracy.
Conditions for an ideal location
No shocks and vibrations
No excessive temperature fluctuations
No direct sunlight
No strong drafts
Select a stable, vibration-free and preferably horizontal location for
your scale.
2.3 Setting up the scale
2.3.1 Levelling the scale
Only a perfectly levelled scale delivers accurate weighing results. The
scale is equipped with a level indicator to make it easier to level the
scale.
f To level the scale, turn the scale’s adjustable feet until the air bubble
in the level indicator is in the inner circle.
METTLER TOLEDO Operator Manual bRite AdvancedOrder number 30209977