Technical data
9
04/12
2 Setting up
2.1 Scope of delivery
f Check if the delivery is complete:
– Scale
– Stainless steel weighing pan
– Power supply, stored in the power supply storage compartment
under the scale
– Operating instructions
– Tower with mounting material, tower version only
2.2 Requirements for scale’s location
The right location is critical to weighing accuracy.
Conditions for an ideal location
• No shocks and vibrations
• No excessive temperature fluctuations
• No direct sunlight
• No strong drafts
• Select a stable, vibration-free and preferably horizontal location for
your scale.
2.3 Setting up the scale
2.3.1 Levelling the scale
Only a perfectly levelled scale delivers accurate weighing results. The
scale is equipped with a level indicator to make it easier to level the
scale.
f To level the scale, turn the scale’s adjustable feet until the air bubble
in the level indicator is in the inner circle.
METTLER TOLEDO Operator Manual bRite AdvancedOrder number 30209977