In all installation work, refer to the appropriate control drawing:
Installation drawing 72227513
Installation drawing for US 72227512
Installation drawing for Canada 72227511
Attaching power cable to the scale
Attaching power cable to the scaleAttaching power cable to the scale
Attaching power cable to the scale
Ensuring admissibility of the peripheral unit
Ensuring admissibility of the peripheral unitEnsuring admissibility of the peripheral unit
Ensuring admissibility of the peripheral unit
If a peripheral unit needs to be connected, its RS232 interface must also be intrinsically safe and
approved to maintain the intrinsic safety.
Ensure that the electrical limiting values of the peripheral match the limiting values printed on the
scale next to the socket.
If this is not the case, install an approved barrier.
Attaching peripheral unit
Attaching peripheral unitAttaching peripheral unit
Attaching peripheral unit
Turn the scale on its side so that the
connection socket (1) on the bottom is
accessible.
Route the blue cable from the Panda
-EX1P Power Supply Unit to the scale
and plug connector into the scale.
Use a grounding cable (cross-section1
mm
2
) to connect the grounding screw (2)
to the nearest connection of the
grounding system of the building
electrical installation.
Turn the scale back to the proper
position.
Connect the cable from the peripheral
device to the 9-pin I/O socket (3) of the
scale. Fix all connectors by tightening
screws.