Installation and Maintenance instructions METTLER TOLEDO BBA242x Paint Mixing Scales with PandaPanda-EX1P Power Supply Unit
Contents 1 2 3 4 5 6 7 8 9 Documentation for the BBA242x Paint Mixing Scale with with PandaPanda-EX1P Power Supply Unit ... 1 Application range ............................................................................................................ 1 Cautionary notes regarding installation ............................................................................ 1 Installation on the scale .................................................................................................. 2 4.
1 Documentation for the BBA242x Paint Mixing Scale with PandaPanda-EX1P Power Supply Unit The BBA242x Paint Mixing Scale with the Panda-EX1P Power Supply Unit is accompanied by the following documentation: • Installation and maintenance instructions • Operating instructions These installation and maintenance instructions apply to all scales with a type designation containing the letters BBA242x.
Installation on the scale ▲ In all installation work, refer to the appropriate control drawing: – Installation drawing 72227513 – Installation drawing for US 72227512 – Installation drawing for Canada 72227511 Attaching power cable to the scale Turn the scale on its side so that the connection socket (1) on the bottom is accessible. Route the blue cable from the Panda -EX1P Power Supply Unit to the scale and plug connector into the scale.
Installation of the PandaPanda-EX1P Power Supply Unit Installation in the safe area Install the power supply in the safe area near a wall socket. If required, the power supply unit can be permanently fixed by using the mounting plates and the 4 screws. Route the power line cable and the cable to the scale so that they are protected against damage. Cables must not be kinked or bent at sharp angles. Plug the power plug into the wall socket of the building electrical installation.
Installation in the hazardous area of zone 2 (only PandaPanda-EX1P/230V) If the power supply unit is installed in zone 2, the power cable must be attached to the screw terminals of the building electrical installation. Requirement An all-pole power disconnecting device must be installed on the supply side of the screw terminals of the building electrical installation.
6 Operation and maintenance Safety measures during operation ▲ Power supply unit and scale may be operated only when the housing is closed. ▲ Warning: Because of the danger of static discharge, discharge, a protective cover may only be used if it is statically uncritical. uncritical Control Inspect the cables, connectors and accessories at regular intervals. Ensure they are free from cracks and other mechanical or chemical damage. If such damage is found, inform METTLER TOLEDO service.
Technical data PandaPanda-EX1P Power Supply Unit Classification 230 V type ATEX II (2) G [EEx ib] II B ATEX II 3 (2)G EEx nA [ib] IIB T4 Mains input 230 V, 50 Hz, 60 mA; Um= 250 V Safety output Uo: 14.3 V, Io: 2.28 A, Po: 4.2 W, Co: 2.28 uF, Lo/Ro: 17.5 H/Ω Classification 120 V type Associated Int. Safe, output Class I, Div 1 / GP C,D Mains input 120 V, 60 Hz, 120 mA; Um= 250 V Safety output Uo: 14.3 V, Io: 2.28 A, Po: 4.2 W, Co: 2.28 uF, Lo/Ro: 17.
FCC and Canadian EMC reglementation This equipment has been tested and found to comply with the limits for a Class A digital device, pursuant to both Part 15 of the FCC Rules and the radio interference regulations of the Canadian Department of Communications. These limits are designed to provide reasonable protection against harmful interference when the equipment is operated in a commercial environment.
Installation Drawing 72227513A METTLER TOLEDO Doc. No. 72227513A Issue date: May.
Installation Drawing 72227512A METTLER TOLEDO Doc. No. 72227512A issue date: May.
Installation Drawing 72227511A METTLER TOLEDO Doc. No. 72227511A issue date: May.
To protect your METTLER TOLEDO product‘ product‘s future: METTLER TOLEDO service assures the quality, measuring accuracy and preservation of value of all METTLER TOLEDO products for years to come. termss of service. Thank you.