8 Specifications
8.1 General Information:
Voltage 110 to 240 Volts AC
Frequency 50/60 Hz
Power Less than 150 VA
IEC Mains fuse 2 of 1A (5 x 20 mm) DA205 Delay (T)
Secondary internal fuse 2 of 2A (5 x 20 mm) M205
Battery Sealed Lead Acid, 12 Volt, 0.8 AH
Certification The Laser Sys
*
Stim 540 complies with the light-emitting and laser
product performance standards set forth in the Code of Federal
Regulations, Title 21 (Food and Drugs), Parts 1040.10 and 1040.11.
Delay from laser emission
indicator illumination to
laser emission
Approximately 2 seconds
Operation time Greater than 30 minutes
International standards:
EN60601-1 1990 Approval and Test Specification Medical Electrical Equipment General
Requirements.
EN60601-2-22 1996 Approval and Test Specification Medical Electrical Equipment Particular
Requirements for Safety - Diagnostic and Therapeutic Laser
Equipment.
Dimensions: 9.4 in. (W) x 4.7 in. (H) x 9.4 in. (D)
Weight: 5.5 lbs.
Operating:
Temperature range: 10 - 40
o
C
Relative humidity: 30% - 90%
Transport and storage:
Temperature range: 0 - 70
o
C
Relative humidity: 10% - 100%
Laser output meter: Infrared Photo Sensor Integrated in to Top Panel, with 20 segment LED
bargraph display. Note: For laser applicatior power verification only.
8.2 Laser Performance:
Output power: Dependent on Applicator (automatically sensed)
Laser diode applicator 80 mW at 785nm
Optional: cluster
applicator
500 mW at 660/950nm
Delivered energy: 0.01 to 99.99 Joules
Operation modes: Continuous and Pulsed
Pulse mode:
Pulse width:
Laser 100 µs nominal
Cluster 50% duty cycle
Pulse frequency:
A) Continuous
B) 10 Hz, 25 Hz, 50 Hz, 100 Hz, 250 Hz, 500 Hz, 1 kHz, 2.5 kHz, 5 kHz
(Pulses per Second)
C) Sweep from 10 Hz to 5 kHz (inc continuous) in 10 seconds (1 second at
each step)