6 Routine Analysis
28 NIRS Vision Tutorial – Operation
6 Routine Analysis
In Routine Analysis you use the models and equations you have built to analyze
unknown samples. All Routine Analysis parameters are brought together in the
Operations Method.
1.8 Setting up an Operations Method
Vision offers a wizard for setting up an operations Method.
Start Operations Method Wizard
1. Click on Mode Data Acquisition
2. Click on Mode Routine Analysis OM Wizard.
3. In the Select Operations Method dialog box, click on [New]
4. In the Select Operation Method Standardization dialog box, set the
following:
Reference Standardization enabled
Click on [OK]
5. In the OM Wizard – OM Name dialog box, enter
OM Name “Tutorial”
Click on [Next]
Select a DCM
1. In the OM Wizard – Select DCM dialog box, select
DCM Diags
Click on [Timer]
2. In the Configure Timer/Reference Frequency dialog box, set the
following:
In the Sample Frequency box:
Sample Collection Interval (sec.) 0
No delay between enabled
In the Time Mode box:
Timed Acquisition disabled
In the Reference Frequency box:
Every x Sample enabled
Frequency x 10
Click on [OK]