Manual
4 Qualitative Model Development
12 NIRS Vision Tutorial – Operation
Include other Products
1. Repeat the Sample Selection Process for all the products in the xdsstart
project.
1.3 Developing an Identification method
An Identification Method is required to differentiate products which differ in
chemical and/or physical properties.
Develop an Identification Method
1. Click on Mode Qualitative Analysis Identify Method
Development
2. File Library Open, select the library xdslib
3. Click on Identify Apply Math.
4. In the Edit Identify Method dialog box, set the following:
Method Correlation in Wavelength Space
Pre Treatment Method 2nd derivative
Region 1
Minimum: 416
Maximum: 1080
Region 2
Minimum: 1120
Maximum: 2484
Threshold type = Match value; value = 0.95
Identify Acceptance Set enabled
Identify Rejection Set disabled
Click on [OK]
5. In the Edit Math Properties dialog box in the Wavelength Regions tab
set the following wavelength regions:
Region 1
Minimum: 416
Maximum: 1080
Region 2
Minimum: 1100
Maximum: 2484
Click on [OK]
6. In the Edit Math Properties dialog box in the 2
nd
Derivative tab set the
following parameters:
Segment: 10
Gap: 0