User guide

▪▪▪▪▪▪▪
21
3.3 Weight
System Type Pounds Kilograms
Basic System: 93 lbs 41.9
With Vortec Cooler: 106 lbs 47.7
With Purge Control and Air Conditioning: 135 lbs 60.8
3.4 Operating Temperature Range
System Type Fahrenheit Centigrade
Basic System: 32-95 0-35
With Vortec Cooler: 32-104 0-40
With Purge Control and Air Conditioning: 32-125 0-51
With Air Conditioning 32-125 0-51
3.5 Operating Humidity Range
System Type % Relative Humidity
All Systems: 10-90 % non-condensing
3.6 Power Requirements
AC Mains Power: 100/240 VAC Selectable, 50/60 Hz, 2A / 1.5A
Power Consumption: 100W Typical, Heater adds up to 150W, Fans add 20W
With Air Conditioning (optional) 5A / 2.7A
Air Consumption:
System Type Air Requirements Fitting Type
Vortec Cooler:
US: 25 SCFM @80 PSI
Metric: 707.9 standard liters/min
@ 5.625 kg/sq. cm
¼” FPT
Purge:
US: 80-120 PSI, set at 60 PSI,
12 SCFM Rapid Exchange
Metric: 5.625 8.44 kg/sq. cm,
set at 4.22 kg/sq. cm,
339.8 standard liters/minute
Rapid Exchange
¼” FPT