3 Instrument Specifications
The XDS Process Analytics Instrument uses Near Infrared (NIR) energy to determine chemical and
physical characteristics of the sample. The sample is illuminated with white light through a fiber optic
bundle. The light is returned through a separate fiber bundle path, and the returning signal is broken
down into visible and NIR wavelengths for spectroscopic analysis. The spectroscopic data is processed
using a PC, and results are output in real time. This permits on-the-fly adjustment of the process to
maintain optimum characteristics.
Instrument Specifications are as follows:
3.1 Enclosure Options for Specific Environments
• Class I Div. I, group B, C and D, Class I Zone 1
X Purge withVortec Cooler; Fiber Optic Communication Modems
• Class I Div. 2
X Purge with Air Conditioner; Fiber Optic Communication Modems
• CENELEC-Approved
X Purge withVortec Cooler; Fiber Optic Communication Modems
• Non-Purged Enclosure Options:
• Fan Cooled, for environments 0-35 degrees C (32-95º F)
• Air-conditioned, for environments 0-51 degrees C (32-125º F)
• Vortec Cooled, for environments 0-40 degrees C (32-104º F)
• Temperature Probes are offered for process monitoring with all models
3.2 Dimensions
Basic System:
• 19.25” (489mm) wide with heat sink and fan assembly
• 24” (609.6mm) high
• 14” (355.6mm) deep
• There is a 3.25” (82.6mm) heat sink and fan on the right side of the enclosure.
Overall Maximum Size with Purge System, fiber hookup area, and other options:
• 30.56” (776.3mm) wide
• 30.45” (773.5mm) high
• 14.06” (357.2mm) deep
Access room should be left on all sides for service, and in front for the door opening.