User guide
6 Devices
IC Net 2.3
80
<Add> Add new program instruction.
<Delete> Delete selected program instruction.
<Verify> Test the time program (error messages are
displayed if program is wrong).
List of program instructions
The following program instructions can be added to the time pro-
gram on the Program subpage:
Instruction Parameter entry Meaning
Remote 0, 1, *
Set remote output lines 1...8 to
the desired values. For entry of
the first value, enter 1, 0 or *. For
entry of the other values, move
the cursor in front of the value to
be changed and enter 1, 0 or *.
Range 100, 200, 5000 µS/cm
1, 2, 5, 10 mS/cm
Set measuring range to the se-
lected value.
FS 0.05 µS/cm ...
10 mS/cm
Set full scale range to the se-
lected value (depends on Range
setting).
Zero on, off Switch on or off the auto-zero
function.
Polarity +, – Reset the polarity of the output
signal.
Mark Trigger marking signal.
ValveA fill, inject Switch injection valve A to "inject"
or "fill" position.
ValveB fill, inject Switch injection valve B to "inject"
or "fill" position.
Suppressor Switch suppressor module to the
next position.
StartDA Start data acquisition.