User guide
6.4 819 IC Detector
IC Net 2.3
77
Damping Display of electronic damping of the analog
output.
Remote lines Display value of remote line settings 1…8.
Zero <ON>
Triggers the auto-zero function.
Zero <OFF> Switch off the auto-zero function.
Method parameter
The Method parameter tab of the 819 IC Detector window contains
the method startup values for the instrument. These values are
sent and applied to the 819 IC Detector by Start determination and
Startup hardware, or if the values are sent manually with <SET>.
Range Measuring range in µS/cm
Selection:
100, 200, 500 µS/cm and
1, 2, 5, 10 mS/cm
Full scale The full scale range (operating range) sets
the desired sensitivity for the analog output.
The possible values of the full scale range
in µS/cm depend on the preset measuring
Range.
Ion Combination of polarity and temperature
coefficient:
Anion polarity = +, temperature coeff. = 2.5
Cation
polarity = –, temperature coeff. = 1.5