56
Events configuration
On the Events configuration subtab user-defined event commands
can be defined, which can be inserted into a time program after
being activated with <Activate>.
Name (1
st
column) User-definable name of the event com-
mand (e.g. Start_766).
Event command (2
nd
column)
Setting the event output lines 1…7.
Selection:
0 (line off, inactive, open)
1 (line on, active, 0 V)
p (pulse, pulse length 150 ms)
* (leave line in current status)
For entry of the first value, enter 1, 0, p or *.
For entry of the other values, move the cur-
sor in front of the value to be changed and
enter 1, 0, p or *.
<Add> Add new remote command.
<Delete> Delete selected remote command.
<Activate> Activate the defined remote commands for
insertion into the time program.