On the Program subpage, program steps including time, program
instruction and parameter can be entered.
First column Time at which program instruction is ap-
plied.
Entry range: 0.0 ... 999.9 min
If no time is entered, the program instruc-
tion is applied together with the last instruc-
tion with time entry.
Second column Program instruction (see below).
In addition to these predefined instructions,
user-defined remote commands can be en-
tered if activated on the Events configuration
tab.
Third column Parameter for program instruction (see be-
low).
ENABLED Enable program start (a disabled program
is not started).
<Add> Add new program instruction.
<Delete> Delete selected program instruction.
<Verify> Test the time program (error messages are
displayed if program is wrong).
List of program instructions
The following program instructions can be added to the time pro-
gram on the
Program subpage:
Instruction Parameter entry Meaning
Events 0, 1, p, *
Set event output lines 1...7 to the
desired values. For entry of the first
value, enter 1, 0, p or *. For entry of
the other values, move the cursor in
front of the value to be changed
and enter 1, 0, p or *.
Program END, RESET The END flag can be used to end a
program, especially if the program
time should be longer than the
chromatogram duration. Additional
steps after this flag are not allowed.
The RESET flag is used to reset the
parameters to the system startup
values.