User guide

5.2 762 IC Interface
IC Net 2.3
47
The Program tab contains the two following subpages:
Program Main time program with all program steps.
Events configuration Possibility for creation of user-defined re-
mote commands.
Program
On the Program subpage, program steps including time, program
instruction and parameter can be entered.
First column Time at which program instruction is ap-
plied.
Entry range: 0.0 ... 999.9 min
If no time is entered, the program instruc-
tion is applied together with the last instruc-
tion with time entry.
Second column Program instruction (see below).
In addition to these predefined instructions,
user-defined remote commands can be en-
tered if activated on the Events configuration
tab.
Third column Parameter for program instruction (see be-
low).
ENABLED Enable program start (a disabled program
is not started).
<Add> Add new program instruction.
<Delete> Delete selected program instruction.
<Verify> Test the time program (error messages are
displayed if program is wrong).
List of program instructions
The following program instructions can be added to the time pro-
gram on the Program subpage:
Instruction Parameter entry Meaning
Events 0, 1, p, *
Set event output lines 1...7 to the
desired values. For entry of the first
value, enter 1, 0, p or *. For entry of
the other values, move the cursor in
front of the value to be changed
and enter 1, 0, p or *.