User guide

4.3 System functions
IC Net 2.3
33
If the Auto restart option is enabled or if a sample
queue is active, a new determination is started immedi-
ately after the running determination is stopped. If you
want to avoid this, disable the Auto restart option or
Pause the sample queue.
SYSTEM / Control / Stop data acquisition
Stop data acquisition of the running determination immediately
and save the recorded chromatogram automatically if the Save
chromatogram after the run
option on the Passport / Processing tab
is enabled. The time programs of the running determination are
continued normally.
4.3.4 Options for determinations
SYSTEM / Control / Auto restart
If this option is enabled, a new determination is started automati-
cally using the current system after the preceding determination
has been finished normally or stopped manually.
The Auto restart option is disabled if determinations are made
with an active sample queue. This option allows to make an infi-
nite batch cycle with the current system
SYSTEM / Control / Verify sample
If this option is enabled, the Edit sample description window is
opened automatically at the start of each determination for entry
of the following sample information:
Ident User defined identifier (title) for the chroma-
togram to be displayed in the title bar of the