User guide

3.4 Global settings
IC Net 2.3
25
Enable Always, if you use «IC Net» together with an Online IC.
In This case Method/Method setup/Export/Send data to Online has
to be checked too, for details see the Online IC instructions
for use.
User comments
Changes on methods, systems, chroma-
tograms,… :
Require Comment on changes
The user has to comment a change.
21 CFR Part 11 On Switch on/off global 21 CFR Part 11 set-
tings. Is automatically activated and cannot
be altered, if IC Net Compliant is installed.
The following parameter settings are auto-
matically set:
Data file overwrite = Never
If method changed = Ask to save
If disk method is newer = Ask to overwrite
User comments = Require Comment
on changes enabled
GLP On = enabled
GLP On Switch on/off global GLP settings. The fol-
lowing parameter settings are automatically
set:
Data file overwrite = Never
If method changed = Ask to save
If disk method is newer = Ask to overwrite
System face
Stacked recorder icons
If this option is enabled, only one recorder
icon with different tabs appears in the sys-
tem window if several data recorders are
installed.
OK, Apply buttons mean
Save object to file If this option is enabled, the <Apply> button
in system settings windows is replaced by
the
<Save> button. The system settings are
saved if the <Save> or <OK> button is
clicked.
Chromatogram units Units for chromatograms:
Flow Unit for flow rate: µL/min, mL/min
Pressure
Unit for pressure: MPa, psi, bar, atm
Print via print spooler
Switch on/off printing via print spooler.
Switch off this option if problems with print-
ing occur.