User guide

6 Devices
IC Net 2.3
258
6.28.4 Hardware settings
Selection of the Hardware menu item using the right mouse button
opens the hardware settings window consisting of the four tabs
Outputs, Alarm stops, System settings and Detector settings.
Outputs
The Outputs tab of the hardware settings window defines remote
output signals to be set automatically if specific events occur.
Event Events for automatic remote signal output:
Inject
Switching of the injection valve to the "IN-
JECT" position.
Fill
Switching of the injection valve to the "FILL"
position.
Pump stop (P max/min)
Pump stopped because pressure limits are
exceeded.
Hardware error
Hardware error detected at the 844 UV/VIS
Compact IC (high-pressure pump, injection
valve, or suppressor not working correctly).
Alarm leak detector
Leak detector has detected solvent in the
instruments interior.
PowerOn values
Display of the power on startup values for
remote output lines set on the Hardware tab.
Pulse length Length of a pulse in ms.
Alarm stops
The Alarm stops tab of the hardware settings window defines the
events for which the 844 UV/VIS Compact IC is stopped immedi-
ately. At an alarm stop, high-pressure pump and peristaltic pump