User guide
6.26 838 Advanced Sample Processor
IC Net 2.3
227
List of program instructions
The following program instructions can be added to the time pro-
gram on the Program tab:
Instruction Parameter entry Meaning
Move sample, sample+1,
ext.#, +rotate, -
rotate, +swing, -
swing, 1...n
Move rack to place the specified
vessel at the needle (see 'Auto-
sampler' in this chapter for an ex-
planation of the positions).
Lift shift, 0, 1...125 mm Raise or lower the lift to the speci-
fied position.
Pump on (infinite), off,
1...999 s
Switch on/off peristaltic pump at
the tower.
Pump
speed
-15, …, -1, 1, …, 15 Change the speed of the peristaltic
pump.
Valve fill, inject Switch injection valve to position
"fill" or "inject"
Scan "scan command" Scan remote interface until the
specified signal is received. In ad-
dition to the normal scan com-
mands (details see 838 Instruc-
tion for Use) special commands
defined on the Configuration tab
can be used.
Ctrl "control
command"
Set the specified remote output
signal. In addition to the normal
control commands (details see 838
Instruction for Use) special
commands defined on the
Configu-
ration
tab can be used.
Wait 1...9999 s Waiting time.
StartDA Start data acquisition.
Dosino
MSB1..3
Injection vol. Dose Volume defined in the EDIT
SAMPLE DESCRIPTION
window or in
the sample queue.
-Injection vol. Aspirate Volume defined in the
EDIT SAMPLE DESCRIPTION window
or in the sample queue.
Concentr. vol. Doses Concentr. vol..
-Concentr. vol. Aspirate Concentr. vol..
The Concentr. vol. is calculated
from values given for Mixing vol-
ume (MV) on the tab Configura-
tion
/
General and the
p
arameter Di-