212
tor P5 AUXILLIARIES. The time program
for each control line is arranged on a sepa-
rate sheet.
AUX-1,…AUX-4 Sheet for the time program of the corre-
sponding control line. For each control line
four events to set the contact closure to
normally open (Aux-X OFF) and four
events to set the contact closure to nor-
mally closed (Aux-X ON) can be pro-
grammed.
Current Displays the actual settings of the Timebase
program
.
Method values Check box to enable/disable the corre-
sponding event.
Disabled
Enabled A combo box appears to select
a time for the corresponding event.
Entry range: 0:00:00…9:59:59 h:mm:ss
<Set> Set parameter values selected under